![NanoString Technologies nCounter Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/nanostring-technologies/ncounter/ncounter_user-manual_705140009.webp)
2017-07 MAN-C0034-06
|
9
NanoString Technologies
®
USER MANUAL
F. Additional Precautions
• The Digital Analyzer is a Class 1 laser product, and the instrument contains an internal Class 2 laser barcode reader. Class 2 laser
radiation exposure is possible while the Digital Analyzer cover is opened. Do not stare into the barcode reader laser beam.
• The Prep Station contains high voltage and heater modules indicated by symbols on the deck. Avoid contacting the heater module
and electrodes. The instrument is equipped with a safety interlock that will prevent voltage from being applied while the instrument
door is open.
• Do not use this device in close proximity to sources of strong electromagnetic radiation or vibration, which may interfere with proper
operation.
• Do not attempt to install, move, or disassemble the instruments.
• Do not override the door sensors (a pinch hazard exists).
• Ensure that all consumables are properly inserted into the system prior to starting a procedure.
• Operate the system using only NanoString nCounter test kits in accordance with their Intended Use.
• Wear gloves while operating or performing user maintenance on the instruments.
• Do not attempt to wash the Prep Station’s electrodes or allow water or other solvents to come into contact with the electrodes.
• Do not attempt to wash either instrument’s touchscreen or allow water or other solvents to come into contact with the touchscreens.
G. Caution Symbols
BIOHAZARD:
Potential hazard from a biological source exists. If you choose to use biohazardous materials on your Prep
Station, the instrument may become contaminated with biohazardous materials. Please affix the proper warning labeling
to your Prep Station if you use biohazardous inputs. Be careful not to touch this area without gloves or other personal
protective equipment.
ELECTRICAL HAZARD
CAUTION HOT SURFACE
H. Environmental Requirements
• Temperature: 18–28°C
• Humidity: < 80% relative humidity (non-condensing)