15) Use the UP key to scroll to “AND”. This makes the
Set Point only true when both threshold conditions
have been met.
Use “OR” for Set Points that trigger when either threshold condition
has been met.
16) Push ENTER. The Operator Relationship is stored.
A dual threshold Set Point is now programmed.
To add functionality to the Set Point, continue on
to any or all of the following procedures or push
the EXIT (SETUP) key to return to normal scale
operation.
1
DEADZONE
0
2
NEW DZ
3
5
4
DEADZONE
1) Use the UP key to scroll to “DEADZONE” or “PRE
ACT” or “POST ACT”.
In this example we’ll add a dead zone of 5 lb.
2) Push ENTER. The current value for the parameter
is displayed, and the message reads “NEW DZ” (or
“NEW PRE” or “ NEW POST”).
3) Use the numeric keys to type in the value required.
4) Push ENTER to store the value.
The value modifier entry is complete. To add
functionality to the Set Point, continue on to any or
all of the following procedures or push the EXIT
(SETUP) key to return to normal scale operation.
To add Preact, Postact or Dead Zone to the Set Point value(s)
To assign a Set Point to a specific CellScale Input or Math Channel
This procedure requires knowledge of the CellScale Channel List. In single channel systems the input source is always 1 which is the default. In that case this
procedure can be skipped. If, however, a Set Point is assigned to a second or higher input or math channel, this procedure is necessary.
For this example, we’ll assume a two entry Channel List, and we’ll assign this Set Point to channel list entry #2.
1
INP SRC
1
2
NEW SRC
3
2
4
INP SRC
1) Use the UP key to scroll to “INP SRC”.
2) Push ENTER. The current input source is displayed,
and the message reads “NEW SRC”.
3) Use the numeric keys to type in the Channel List
position number.
4) Push ENTER to store the value.
The input source entry is complete. To add function-
ality to the Set Point, continue on to any or all of the
following procedures or push the EXIT (SETUP)
key to return to normal scale operation.
1
OUT DEST
1
2
NEW DEST
3
2
4
OUT DEST
1) Use the UP key to scroll to “OUT DEST”.
2) Push ENTER. The current CellScale output destina-
tion is displayed, and the message reads “NEW
DEST”.
3) Use the numeric keys to type in the Set Point Output
List position number.
4) Push ENTER to store the value.
The output destination entry is complete. To add
functionality to the Set Point, continue on to any or
all of the following procedures or push the EXIT
(SETUP) key to return to normal scale operation.
To Setup the CellScale Set Point Output
This parameter affects the CellScale only and has no effect on this or any other connected 3750CS