Page 16 MSI-3750CS RF Weighmeter • User Guide
3750CS
C
ELL
S
CALE RF DIGITAL WEIGHT INDICATOR
®
Push ZERO. The weight reading must be stable
within ±1 division for the zero function to work.
The display temporarily reads “ZEROED” and the
digits display 0. The backup memory stores the zero
reading, and can restore it even if power fails.
Final: The numeric digits display “0” (or 0.0 or
0.00, etc.).
ZEROED
1) Push the SC INDEX key (F3 in this example). The
message “SC INDEX” appears, and the numeric
digits indicate the current SC Index # (1-32).
2) Within three seconds, push either the SC INDEX
key again or the UP scroll key. The Scale Channel
will index to the next channel in the list.
To go to the previous channel in the list, push the DOWN scroll key
within 3 seconds of the SC INDEX key. (This won’t work if you used
F2 for the SC INDEX key)
3) Either wait three seconds for normal display func-
tion to return, or you can push ENTER or EXIT
to hasten return to standard scale function, now
operating on a new scale channel.
.01
1
SC INDEX
.02
2
SC INDEX
3
or wait 3 seconds.
Multiplexer and provide individual wheel loads. Index position 5 shows front axle load, index 6 shows rear axle
load, and index 7 shows total weight. The CellScale allows the user to name each channel and the name can
appear on the message (lower) display of the 3750CS. When dealing with multiple channels it is recommended
that you program the lower display mode to either display the scale channel number (LC NUMBR), the Scan
List Index number (SC INDEX), or the channel name (LC NAME).
See the “METER SETUP” section 7 to configure the lower display mode.
For more information on Scan Lists and dealing with multiple channels, please refer to the CellScale manual.
TO ACCESS MULTIPLE CHANNELS
As mentioned above, the 3750CS will access any channels setup in the CellScale. It does this with the “SC
INDEX” function. The SC INDEX steps from channel to channel as dictated by the Scan list stored in the
CellScale. If the CellScale has only one scale channel defined, this procedure is not necessary. This is always
the case for MSI-6260CS Crane Scales and other single load cell systems.
For systems with multiple scales attached to any given CellScale the setup procedure is:
1) Program the CellScale Scan List as required (see the CellScale Manual).
2) Program a function key for “SC INDEX” (see section 4 – FUNCTION KEYS).
3) Optional but recommended so you can readily identify which scale you are monitoring and controlling
– Program the lower display mode for “LC NUMBER”, “SC INDEX”, or “LC NAME” (see section
7 – METER SETUP).
To Change Scale Channel
This example assumes F3 has been setup for “SC INDEX”.
ZERO
Sets the zero reading of the scale. Use the zero key to take out small deviations in zero when the scale is
unloaded. (See “TARE SET” for zeroing (Taring) package or pallet weights)
To Zero
Rules for Use:
1) Works in GROSS mode or NET mode. Zeroing while in Net mode will zero the gross weight causing the
display to show the negative Tare value.
2) The scale must be stable. The scale will not zero if the motion detect annunciator is on.
3) The scale will accept a zero setting over the full Range of the scale (NTEP and other Legal-for-trade models
may have a limited zero range). Zero settings above 4% of full scale will subtract from the overall capacity
of the scale. For example if you zero out 100 lb. on a 1000 lb. scale the overall capacity of the scale will
reduce to 900 lb. plus the allowed over-range amount.