Chapter Nine: RS-232 Communications
Command Messages
277
Read Only Data Messages (Continued)
Message Description
Message Format
@ command
category
command
number
Parameter
Data
Optional
checksum
Relay Status
@
6
0
A
A, B, C, or D
?
CR c1
c2
Alarm Status
@
6
0
B
A, B, C, or D
?
CR c1
c2
Current Control Mode @
6
0
C
Any character
?
CR c1
c2
Current Recipe
@
6
0
D
Any character
?
CR c1
c2
Current PID Set Point @
6
0
E
Any character
?
CR c1
c2
PCS Output
@
6
0
F
Any character
?
CR c1
c2
Table 78: Read Only Data Messages
Response Format for Read Only Data Messages
Message
Response*
Description
Channel Condition
@608X:<status><pres>
where x is 1 through 6
<pres> is an empty field unless <status> is A or
B, then <pres> is an ASCII string
<status> is:
A = On
B = low power degas
C = Underranged
D = Overranged
E = Manually off
F = Auto off
G = High power degas
H = Initializing
I = Zeroing
J = Bad sensor
K = Disconnected
L = Channel not
installed
Relay Status
@60A:<para><status>
where <para>
A, B, C, or D
<status> is
E = Enabled
D = Disabled
T = Tripped
Table 79: Response Format for Read Only Data Messages
(Continued on next page)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com