How To Set Alarm Trip Points
Chapter Seven: Operation in Setup Mode
188
How To Set Alarm Trip Points
MAIN DISPLAY
The trip point value for the selected
alarm is shown in this display.
CLOSE
TORR
SETUP
TRIP POINT
LEFT DISPLAY
CENTER DISPLAY
RIGHT DISPLAY
The letter of the alarm
which is selected for
editing (A through D)
is shown here.
In this example, alarm
B is selected.
This is the channel number
for the gauge which is
associated with the selected
alarm. The uppercase A
indicates the untripped state
of the relay is actuated.
This indicates whether the
selected alarm is ON (enabled)
or OFF (disabled) as shown in
this example.
Figure 54: Display for Setting Alarm Trip Points
The trip point is the pressure at which an alarm is tripped. Alarms A and B are standard, but
alarms C and D are available only if the optional Auxiliary Output board is installed.
Note
Be sure the input channel to the alarm is powered on. You will not be
able to configure the alarm unless the input channel is powered on.
1. Repeatedly press the [
DISPLAY MODE
] key until the 146 unit is in Setup Mode.
The system responds by scrolling through modes.
2. Use the arrow keys to scroll to the
TRIP POINT
legend.
The system responds by scrolling through legends.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com