Code 11x: How To Turn Auto Zeroing ON or OFF
Chapter Seven: Operation in Setup Mode
228
Code 11x: How To Turn Auto Zeroing ON or OFF
SETUP
CODE
RIGHT DISPLAY
LEFT DISPLAY
The code number is displayed
and edited here.
In this example, the
code number is 11x,
where x represents
the channel number for
the gauge to be zeroed.
This display toggles between
on (enable the auto zero feature), and
oFF (disable the auto zero feature), as
in this example.
MAIN DISPLAY
This is the channel number of the reference channel,
by which the selected channel (left display) is zeroed.
Figure 74: Code 11x Display
The x in the code number represents a channel number. When the Auto Zero feature is set up
and enabled, one channel (defined as the reference channel) automatically zeros a second channel
(defined as the selected channel). The selected channel is continually zeroed as long as the
reference channel’s pressure is below the selected channel’s minimum resolution. The zeroing
operation is suspended whenever the reference channel’s condition is not measuring, and during
the first 10 seconds after the reference channel is turned on.
The Auto Zero function overrides any user-defined zero previously set. If a new user-defined
zero is then implemented, it overrides the effects of the Auto Zero function. If the Auto Zero
feature is disabled, the 146 unit uses the last zero defined (by Auto Zero or user-defined, and via
the front panel, RS-232, or rear panel remote zero).