Chapter Seven: Operation in Setup Mode
How To Perform Sensor Calibration/MFC Setup
193
5. Enter the sensor resolution in the center display.
Sensor resolution is the smallest pressure that can be displayed above or below zero. The
sensor resolution range for a Type 120 gauge or linear capacitance manometer is
±7 decades.
The default resolution for all capacitance manometers is 10-
1
.
A. If the display is correct, then press the [
ENTER
] key.
B. If the display is incorrect, enter the correct sensor resolution, then press the [
ENTER
]
key.
The system responds by accepting the sensor resolution.