Chapter Three: Overview
Inside the Type 146 Unit
73
9. Replace the bottom cover by inserting it into the grooves provided, and firmly sliding it
towards the front of the unit. Replace the seven (7) screws that fasten the bottom cover.
The center screw on the bottom cover is slightly off center to avoid interfering with a
connector on a board in slot 5. Therefore, the cover must be in the correct orientation for
the screw holes to line up properly.
10. Carefully turn the unit right-side up.
11. Replace the top cover by inserting it into the grooves provided, and firmly sliding it
towards the front of the unit. Replace the seven (7) screws that fasten the top cover.
The center screw on the bottom cover is slightly off center to avoid interfering with a
connector on a board in slot 5. Therefore, the cover must be in the correct orientation for
the screw holes to line up properly.
12. Replace the four (4) Phillips screws on the outside corners of the rear panel and the seven
(7) screws on the top cover.
13. Replace the four (4) screws on the rear panel.
14. Plug in the power cord.