How To Configure an Analog Output
Chapter Seven: Operation in Setup Mode
210
3. Enter the number for the analog output to be modified.
A. If the left display is correct, press the [
ENTER
] key.
B. If the left display is not correct, enter the correct analog output number (1 through 3),
then press the [
ENTER
] key.
The system responds by accepting the analog output number, and causing the main
display to blink.
4. Enter the pressure value corresponding to full scale voltage OR the voltage for set point
(0 to 10 Volts), if that is the input method to be used.
A. If the main display is correct, press the [
ENTER
] key.
B. If the main display is not correct, enter the correct pressure value, then press the
[
ENTER
] key.
Note
The pressure units remain the same, even if the value entered is intended
to be a fixed set point voltage.
5. Select the output scaling method to be used.
A. If the right display is correct, press the [
ENTER
] key.
B. If the right display is not correct, use the arrow keys to select the output scaling
method, then press the [
ENTER
] key.
If ln or lo1-lo9 is selected, the system responds by accepting the output scaling method,
and causing the center display to blink.
If SP is selected, the center display goes blank, the units in the main display change to
VOLTS
, and the left display blinks. If the value entered in the main display is greater
than 10, the system defaults to 5 Volts.
6. Select the input channel (1 through 4, 5 for dual channel 1, or 6 for dual channel 2)
which supplies the input signal for scaling.
A. If the center display is correct, press the [
ENTER
] key.
B. If the center display is not correct, enter the correct input channel (1 through 4, 5 for
dual channel 1, or 6 for dual channel 2), then press the [
ENTER
] key.
The system responds by accepting the input channel number.
Note
The same channel may be used for more than one analog output.