7-2
NOTE
If the message "SpO
2
Low Perf." or "SpO
2
Weak Pulse" displays, check sensor
application, re-apply or remove the sensor if necessary, to obtain a better
signal.
A function tester or SpO
2
simulator can be used to verify the sensor functions.
A functional tester or SpO
2
simulator cannot be used to assess the accuracy of a
SpO
2
module or a SpO
2
sensor.
A functional tester or SpO
2
simulator can be used to determine the pulse rate
accuracy.
7.2
Safety
WARNING
Use only SpO
2
sensors specified in this manual. Follow the SpO
2
sensor’s
instructions for use and adhere to all warnings and cautions.
When a trend toward patient deoxygenation is indicated, blood samples should
be analyzed by a laboratory co-oximeter to completely understand the patient’s
condition.
Do not use SpO
2
sensors during magnetic resonance imaging (MRI). Induced
current could potentially cause burns. The sensor may affect the MRI image,
and the MRI unit may affect the accuracy of the oximetry measurements.
Prolonged continuous monitoring may increase the risk of undesirable changes
in skin characteristics, such as irritation, reddening, blistering or burns. Inspect
the sensor site every two hours and move the sensor if the skin quality changes.
Change the application site every four hours. For neonates, or patients with
poor peripheral blood circulation or sensitive skin, inspect the sensor site more
frequently.
7.3
Identifying SpO
2
Module
To identify which SpO
2
module is incorporated into your monitor, see the company logo
located at the side panel. The color of the cable connector matches the company as
shown below:
Mindray SpO
2
module: a blue connector without logo.
Masimo SpO
2
module: a purple connector with a logo of Masimo SET.
Nellcor SpO
2
module: a grey connector with a logo of Nellcor.
The connectors for these three SpO
2
sensors are mutually exclusive.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...