C-4
Temp unit
°C
Yes
No
C.1.5
CO
2
Name
Default Config
Saved at Power
Failure
Affected by
Defaults
Parameter-RR
Alarm On/Off
On
Yes Yes
Alarm Priority
Med
RR High
30
RR Low
8
Parameter-EtCO
2
Alarm On/Off
On
Yes Yes
Alarm Priority
Med
EtCO
2
High
50
EtCO
2
Low
25
Parameter-FiO
2
Alarm On/Off
On
Yes Yes
Alarm Priority
Med
FiCO
2
High
4
Apnea Delay
20 s
Yes
Yes
Operating Mode
Measure
Yes
No
BTPS Compensate
Off
Yes Yes
Scale 50
Sweep 6.25
mm/s
Wave Type
Draw
O
2
Compensate
21
N
2
O Compensate
0
AA Compensate
0
Auto Standby
60 min