6-5
WARNING
Make sure that the alarm limits settings are appropriate for your patient before
monitoring.
Setting alarm limits to extreme values may cause the alarm system to become
ineffective. For example, High oxygen levels may predispose a premature infant
to retrolental fibroplasia. If this is a consideration do NOT set the high alarm
limit to 100%, which is equivalent to switching the alarm off.
6.5
Selecting Alarm Properties
6.5.1
Changing the Alarm Volume
Select [
Main
]
→
[
General Setup >>
]
→
[
Alarm Volume
].
The alarm volume range is between X to 10. X is the minimum volume, which depends on
the setting of minimum alarm volume (Refer to
6.5.2 Setting the Minimum Alarm
Volume
), and 10 is the maximum volume.
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears
on the screen.
6.5.2
Setting the Minimum Alarm Volume
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3. Select
[
Minimum Alarm Volume
] and then select the appropriate settings.
The minimum alarm volume refers to the minimum value you can set for the alarm
volume, which is not affected by user or factory default configurations.
6.5.3
Setting the Interval between Alarm Sounds
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access [
User Settings
] menu.
2. Select
[
Alarm Setup >>
] to access [
Alarm Setup
] menu.
3. Select
[
High Alarm Interval (s)
], [
Med Alarm Interval (s)
] and [
Low Alarm Interval
(s)
] in turn and then select the appropriate settings.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...