EMC and Radio Regulatory Compliance
B.1
EMC
The device meets the requirements of IEC 60601-1-2.
NOTE
Using accessories, transducers and cables other than those specified may result
in increased electromagnetic emission or decreased electromagnetic immunity
of the patient monitoring equipment.
The device or its components should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the device or its
components should be observed to verify normal operation in the
configuration in which it will be used.
The device needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the
requirements of CISPR.
When the inputted signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of
this monitor.
Other devices that have RF transmitter or source may affect this device (e.g. cell
phones, PADs, PCs with wireless function).
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...