14-6
2. Select
[Import certificates >>
] or [
Delete certificates >>
].
14.6
Setting the Multicast Parameters
Whether the equipment is presented by broadcast or multicast is defined before the
equipment leaves the factory. If [
Multicast
] is selected, you need to set the multicast
parameters.
To do so,
1 Select
[
Main
]
→
[
Maintenance>>
]
→
[
User Settings >>
]
→
enter the required
password
→
Select [
Ok
]
→
[
Network >>
]
→
[
Multicast Setup >>
].
2. Set
[
Multicast Addr
] and [
TTL
].
14.7
DIAP Communication Setup
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access user settings menu.
2. Select
[
DIAP Communication>>
] to access [
DIAP Communication Setup
] menu.
3.
Set up the baud rate.
4. Select
[
Ok
] to exit the menu.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...