15-3
will significantly shorten the life expectancy of a battery. Do not store the
battery at a temperature beyond -20°C- 60°C.
WARNING
Keep the battery out of children’s reach.
Use only specified batteries.
If the battery shows signs of damage or signs of leakage, replace it immediately.
Do not use a faulty battery in the monitor.
15.5
Battery Maintenance
15.5.1
Conditioning a Battery
A battery should be conditioned before it is used for the first time. A battery conditioning
cycle is one uninterrupted charge of the battery, followed by an uninterrupted battery
discharge and charge. Batteries should be conditioned regularly to maintain their useful
life.
To condition a battery, follow this procedure:
1.
Disconnect the monitor from the patient and stop all monitoring or measuring.
2.
Insert the battery in need of conditioning into the battery slot of the monitor.
3.
Connect the monitor to AC power. Allow the battery to be charged uninterrupted till
the battery is full and the battery indicator is off.
4.
Remove AC power and allow the monitor to run from the battery until it shuts off.
5.
Again connect the monitor to AC power. Allow the battery to be charged
uninterruptedly till the battery is full and the battery indicator is off.
NOTE
The actual battery capacity will decrease over time with use of batteries. For old
batteries, the full capacity battery symbol does not indicate the capacity and
operating time of this battery can still fulfill battery specifications in the
operator’s manual. Please replace the battery if its operating time is
significantly lower than the specified time.
15.5.2
Checking a Battery
The performance of a rechargeable battery may deteriorate over time. The battery
performance test must be performed once a year, before monitor repairs, or whenever the
battery is suspected as being the source of the problems.
To check the performance of a battery, follow this procedure:
1.
Disconnect the monitor from the patient and stop all monitoring or measuring.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...