7-3
7.4
Applying the Sensor
WARNING
If the sensor is too tight because the application site is too large or becomes too
large due to edema, excessive pressure for prolonged periods may result in
venous congestion distal from the application site, leading to interstitial edema
and tissue ischemia.
NOTE
Place the SpO
2
sensor so that the light source is against the application site.
Check if the sensor is in normal condition before monitoring. Do not use the
SpO
2
sensor if the package or the sensor is found damaged.
Do not apply the sensor on a limb with an intravenous infusion or arterial
catheter in place.
1.
Select an appropriate sensor according to the module type, patient category and
weight.
2.
Clean the application site. For example, remove colored nail polish.
3.
Apply the sensor to the patient.
4.
Select an appropriate adapter cable according to the connector type and plug this
cable into the SpO
2
connector.
5.
Connect the sensor cable to the adapter cable.
7.5
Changing SpO
2
Settings
7.5.1
Accessing SpO
2
Menu
By selecting the SpO
2
parameter or waveform area, you can access the [
SpO
2
Setup
]
menu.
7.5.2
Adjusting the Desat Alarm
The Desat alarm is a high level alarm notifying you of potentially life threatening drops in
oxygen saturation.
Select [
Alarm Setup >>
] from the [
SpO
2
Setup
] menu. From the pop-up menu, you can
set low alarm limit and alarm switch for [
Desat
]. When the SpO
2
value is below the Desat
alarm limit and Desat alarm switch is set to [
ON
], the message [
SpO
2
Desat
] is displayed.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...