14-3
network from being accessed by unauthorized users.
The total throughput of all the wireless devices connected to the wireless
network should be less than the effective transmitting capability of the wireless
network. The throughput capacity of a single VS-900 is 700 kbps.
Do not connect non-approved devices to the wireless network.
Where the monitor is located, the signal strength of other Wi-Fi devices on the
same channel should be no greater than -85 dBm.
Where the monitor is located, the signal strength of other Wi-Fi devices on
adjacent channels should be no greater than -50 dBm.
The recommended distance between the patient monitor and other non-Wi-Fi
wireless devices, including wireless devices at the frequency of 2.4GHz (e.g.
cellular mobile communication networks, microwave ovens, interphones,
cordless phones and electro-surgical units) is no less than 20 cm.
14.3
Setting Data Send Method
You can choose to send the monitor data to hospital system through eGateway or HL7
protocol.
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
]
→
[
Network>>
].
2. Set
[
Data Send Method
] to [
eGateway Send
] or [
HL7 Direct Send
].
14.3.1
ADT Communication Setup
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
]
→
[
Network >>
]
→
[
ADT Communication>>
].
2.
Set IP address and port, and switch on/off ADT query.
3. Select
[
Ok
] to exit the menu.
14.3.2
EMR Communication Setup
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
]
→
[
Network>>
]
→
[
EMR Communication >>
].
2.
Set the IP address and port.
3. Select
[
Ok
] to exit the menu.
14.3.3
NTP Server Setup
1. Select
[
Main
]
→
[
Maintenance >>
]
→
[
User Settings >>
]
→
Enter the required
password
→
Select [
Ok
] to access User Settings menu.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...