10-4
In Predictive mode, the monitor automatically enters Monitor mode in the following
cases:
Accurate temperature is not reached.
Neither measurement is taken nor is the probe replaced in the probe well in 60
seconds after the probe is withdrawn from the well.
The temperature type automatically changes to Predictive mode when the probe returns
to the probe well.
NOTE
In Predictive mode, temperature probe shall be placed to the measurement site
as soon as probe warmup is complete; otherwise, inaccurate temperature
reading may result.
In Predictive mode, if the probe has a high temperature due to the
environmental temperature or other causes, cool the probe and then measure
the patient's temperature.
10.1.2.5
Taking a Temperature in Monitor Mode
To measure a temperature in the Monitor mode,
1.
Verify that the probe is placed in the probe well.
2.
Verify that the temperature measurement type and site are correct.
3.
Unplug the probe from the probe well and insert it into a cover in the probe cover
pack. Press the probe handle down firmly until the cover engages with the probe.
4.
Place the probe to the measurement site and then start measuring. Refer to Step 4 in
10.1.2.4 Taking a Temperature in Predictive Mode
for how to place a probe.
5.
Withdraw the probe. Press firmly the ejection button on the top of the probe to eject
the probe cover. Replace the probe into the probe well.
NOTE
In Monitor mode, record the measured value prior to taking away the probe
from measurement site. The monitor will automatically stop measuring
temperature after 10 minutes from the start of the measurement.
10.1.3
Disinfecting Temperature Probe
The recommended disinfectants include: ethanol 70%, isopropanol 70%,
glutaraldehyde-type 2% liquid disinfectants.
To disinfect the temperature probe:
1.
Disconnect the temperature probe from Temp connector.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...