D-8
CO
2
Sensor
High Temp
No Low
Ambient temperature
is too high or there is a
module failure.
Lower the operating
temperature.
If the alarm persists, the
CO2 module may fail.
Contact your service
personnel.
CO
2
Zero
Failed
No
Low
The zeroing failed.
Check the CO
2
connections. If the alarm
persists, contact your
service personnel.
CO
2
No
Filterline
Yes Low
The sample line is off
or disconnected.
Make sure that the sample
line is connected.
CO
2
FilterLine
Occluded
No Low
The airway or water
trap is occluded.
Check the airway and
remove the occlusion.
XX
Overrange
No Low
The measured
parameter or
measurement XX
value is not within the
specified range.
Contact your service
personnel.
XX represents a measurement or parameter label.
D.2.5
Recorder Alarm Messages
Alarm
Message
Clearable?
(Yes/No)
Alarm
Level
Cause
Solution
Recorder Init
Error
Yes
Low
An error occurred to
the module, or there is
a problem with the
communication
between the module
and the monitor.
Restart the monitor. If the
problem still exists,
contact your service
personnel.
Recorder
selftest Error
Yes Low
Recorder
Unavailable
Yes Low
Recorder
Comm Error
Yes Low
Recorder Vlt
High
No Low
There is a problem
with the system power
supply. Restart the
monitor.
If the message is
prompted for several
times, contact your service
personnel.
Recorder Vlt
Low
No Low
Recorder
Head Hot
No Low
The recorder has been
working for too long
time.
Stop the recording and
resume the recording till
the recorder’s print head
cools down.
Содержание VS-900
Страница 1: ...VS 900 Vital Signs Monitor Operator s Manual ...
Страница 2: ......
Страница 14: ...8 FOR YOUR NOTES ...
Страница 26: ...2 6 2 3 4 Bottom View 1 Battery compartment door 2 Hole for installing a support 1 2 ...
Страница 28: ...2 8 Trend Screen NIBP List Screen 1 Patient Information System Message Area 10 3 6 8 7 2 1 9 3 5 6 8 7 2 1 5 ...
Страница 42: ...3 10 FOR YOUR NOTES ...
Страница 68: ...6 12 FOR YOUR NOTES ...
Страница 76: ...7 8 FOR YOUR NOTES ...
Страница 78: ...8 2 FOR YOUR NOTES ...
Страница 88: ...9 10 FOR YOUR NOTES ...
Страница 114: ...12 8 FOR YOUR NOTES ...
Страница 139: ...16 11 2 Select Modify Password 3 Enter new password and then select Ok ...
Страница 140: ...16 12 FOR YOUR NOTES ...
Страница 166: ...B 8 FOR YOUR NOTES ...
Страница 191: ......
Страница 192: ...P N 046 004883 00 7 0 ...