6

   Worklight LED ON/OFF

To turn the light on and off, press the   button. 

HOLD Key

Data Hold Function - Freezes the value on the dis-

play. Press the “HOLD” button to freeze the reading. 

The reading will be held regardless of subsequent 

variation in input. HOLD is displayed with the read-

ing. To exit Data Hold mode, press the HOLD button 

again.

CAUTION

The Data Hold readings are re-

leased when the meter enters Sleep 

Mode.

Sleep Mode

The  clamp  meter  is  automatically  powered  off  in 

about  20  min  after  the  last  Rotary  Dial  or  button 

operation. To reset, rotate the Rotary Dial to OFF. 

If the display is still blank when a new Rotary Dial 

setting is selected, replace the batteries.

The  sleep  mode  is  disabled  when  the  MIN/MAX 

function is selected.

The clamp meter does use battery power in sleep 

mode. Be sure to switch the tool to OFF to conserve 

battery power.

Non-Contact Voltage Detection (NCVD)

DANGER

The LED may not be displayed due 

to installation condition of electri-

cal circuit or equipment. Never touch the circuit 

under test to avoid possible danger even if the 

LED for NCVD is not displayed.

Check the functionality of LED on a well-known 

power supply prior to measure-

ment. When the LED doesn’t 

light up, do not make mea-

surement.

NCVD indication is affected by 

external voltage, and how the 

meter is held or placed.

When  the  meter  is  on  in  any 

function,  the  non-contact  volt-

age detector will indicate with a 

Red LED on the display when 

an electric field exceeding 90V is 

detected. Place the edge of the 

jaw  labeled  “Voltage  Detector” 

near the electric field.

Over-flow indication

Any time the input exceeds the measuring range “OL” 

or “-OL” is displayed.

CAUTION

Clamp the jaws around the conduc-

tor under test and press the MIN/

MAX  button  after  an  appropriate  range  is  se

-

lected by Auto-ranging function. ZERO and Hz 

keys  are  disabled  while  MIN/MAX  Function  is 

active.

MIN/MAX Function 

The MIN/MAX function can be used

 

during  measurements  of AC  or  DC 

current, AC or DC voltage, Tempera-

ture (2236-20 only), uA DC (2236-20 

only), and Resistance. The MIN/MAX 

function does not work in the Capaci-

tance or Continuity measurements.

To measure the minimum or maximum 

of the function, set the dial to the ap-

propriate dial position and then press 

the  MIN/MAX  button  to  capture  the 

maximum  reading.  Press  the  MIN/

MAX  button  again  to  capture  the 

minimum reading. 

Press the MIN/MAX button to toggle 

between minimum or maximum. The 

minimum  or  maximum  reading  is 

displayed and held until the MIN/MAX is turned off 

by holding in the MIN/MAX button for 2 seconds or 

changing the Rotary Dial position. 

CAUTION

Auto-Ranging and MIN/MAX button 

are disabled when ZERO function 

is in use. ZERO function operates only in AC 

Current  , DC Current   and Capacitance 

.

ZERO Function

AC Current   / DC Current  :

  When  the  LCD 

doesn’t read ZERO while the

 

Jaws are closed, press the 

ZERO  Button  to  indicate 

ZERO  before  starting  a 

measurement. In this case, 

the  “ZERO”  mark  appears 

on the LCD to indicate the 

ZERO Function is activated.

Capacitance 

Effective 

at 40nF Range only. Press 

the  ZERO  Button  with  the 

Test  leads  opened  before 

measuring capacitances to 

indicate ZERO before start-

ing a measurement.

MAINTENANCE

WARNING

To reduce the risk of injury, always 

remove the batteries from the tool 

before performing any maintenance. Never disas

-

semble the tool. Contact a MILWAUKEE service 

facility for ALL repairs.

Maintaining Tool 

Keep your tool in good repair by adopting a regular 

maintenance program. After one year, it is recom-

mended to return the tool to a MILWAUKEE service 

facility for calibration.

If the tool does not start or operate at full power with 

fully charged batteries, clean the contacts on the bat-

tery door. If the tool still does not work properly, return 

the tool to a MILWAUKEE service facility for repair.

WARNING

To reduce the risk of personal in-

jury and damage, never immerse 

your tool, battery pack or charger in liquid or 

allow a liquid to flow inside them.

Содержание 2236-20

Страница 1: ...injury user must read and understand operator s manual AVERTISSEMENT Afin de r duire le risque de blessures l utilisateur doit lire et bien comprendre le manuel ADVERTENCIAPara reducir el riesgo de l...

Страница 2: ...anch or authorized service station Do not try to replace the batteries if the surface of the instrument is wet Disconnect all the cords and cables from the object under test and power off the instrume...

Страница 3: ...IONS Dial Position Range Resolution Accuracy Voltage DC 60 600V 0 01V 0 1V 1 0 rdg 3dgt Voltage AC 0 0 to 600 0V 0 1V 1 5 rdg 4dgt 50 60Hz 3 5 rdg 5dgt 40 400Hz Current AC 0 to 600 0A Peak 1500A CF 2...

Страница 4: ...o avoid an electrical hazard turn the Rotary Dial to OFF and discon nect the test leads before replacing batteries Loading Changing the Batteries Replace batteries when the Low Battery indicator is di...

Страница 5: ...short circuit the tips of test leads to make the indication zero A buzzer will sound 3 Connect the test leads to both ends of the conductor under test If the resistance under test is 35 10 or 25 or le...

Страница 6: ...MAX Function The MIN MAX function can be used during measurements of AC or DC current AC or DC voltage Tempera ture 2236 20 only uA DC 2236 20 only and Resistance The MIN MAX function does not work i...

Страница 7: ...s provided at the time warranty service is requested ACCEPTANCE OF THE EXCLUSIVE REPAIR AND REPLACEMENT REMEDIES DESCRIBED HEREIN IS A CONDITION OF THE CON TRACT FOR THE PURCHASE OF EVERY MILWAUKEE PR...

Страница 8: ...travailler dans un endroit renfermant des conducteurs sous tension et non prot g s AVERTISSEMENT Ne jamais tenter de prendre une mesure dans des conditions anormales par exemple si le bo tier de l in...

Страница 9: ...Position du cadran Plage R solution Pr cision Tension c c 60 600 V 0 01 V 0 1 V 1 0 rdg 3 dgt Tension c a 0 0 600 0 V 0 1 V 1 5 rdg 4 dgt 50 60 Hz 3 5 rdg 5 dgt 40 400 Hz Courant c a 0 600 0 A Cr te...

Страница 10: ...donn es est d sactiv e Autrement il sera impos sible de prendre la mesure d sir e Afficheur ACL avec r tro clairage L afficheur ACL avec r tro clairage s teint apr s trois minutes d inactivit Appuyer...

Страница 11: ...order le fil d essai rouge la borne positive et le fil d essai noir la borne n gative du circuit mis l essai La lecture s affiche Une con nexion invers e fera s afficher une valeur n gative Mesure de...

Страница 12: ...pince s teint automatiquement environ 20 min utes apr s que le cadran rotatif ou les boutons aient t actionn s pour la derni re fois Pour r initialiser l instrument tourner le cadran rotatif la posit...

Страница 13: ...orrectement renvoyez l outil le chargeur et la batterie un centre de service MILWAUKEE accr dit AVERTISSEMENT Pourminimiserlesrisques de blessures ou de dom mages l outil n immergez jamais l outil la...

Страница 14: ...ous les frais de transport li s ce processus de garantie Exceptions Cette garantie ne s applique pas dans les situations suivantes a Si le produit a t utilis pour une fin autre que celle indiqu e dans...

Страница 15: ...nte No exponga el instrumento a la luz solar directa altas temperaturas humedad ni roc o Para funcionar a una altitud de 2 000 m o menos La temperatura de funcionamiento adecuada oscila entre 10 C y 5...

Страница 16: ...0 0 V 0 1 V 1 5 rdg 4dgt 50 60 Hz 3 5 rdg 5dgt 40 400 Hz Corriente alterna 0 a 600 0 A Pico 1 500 A CF 2 5 600 A CF 3 0 500 A 0 1 A 2 0 rdg 5dgt 50 60 Hz 3 5 rdg 5dgt 40 400 Hz Agregue 2 a CF 2 5 A ag...

Страница 17: ...O cero para ajustar la pantalla 3 Presione el gatillo de abertura de mordazas para abrir las mordazas y sujete con ellas el conductor que se est probando La lectura se muestra en la pantalla NOTA No s...

Страница 18: ...se est probando 5 La lectura se muestra en la pantalla Corr cont Cat No 2236 20 nicamente PELIGRO Para evitar descargas el ctricas Nunca efect e una medici n en un circuito que tenga un voltaje super...

Страница 19: ...MAX para registrar la lectura m xima Presione el bot n MIN MAX de nuevo para registrar la lectura m nima Presione el bot n MIN MAX para alternar entre la lec tura m nima o m xima Se muestra la lectura...

Страница 20: ...el servicio por garant a LA ACEPTACI N DE LOS RECURSOS EXCLUSIVOS DE REPARACI N Y REEMPLAZO DESCRITOS EN ESTE DOCUMENTO ES UNACONDICI N DEL CONTRATO PARALACOMPRADE CADA PRODUCTO MILWAUKEE SI USTED NO...

Отзывы: