80
Micro Motion
®
Model 2200S Transmitters
Testing and Moving to Production
Figure 7-8
ProLink II – mA output trim procedure
Read mA output at
receiving device
Read mA output at
receiving device
Equal?
Enter receiving device
value in
Enter Meas
Read mA output at
receiving device
Read mA output at
receiving device
Equal?
Enter receiving device
value in
Enter Meas
4 (12) mA trim
20 mA trim
Yes
No
Yes
Next
Next
Next
ProLink >
Calibration >
Milliamp 1 Trim
Back
No
Back
Finish
Set mA measurement point