Configuration and Use Manual
117
Troubleshooting
Op
eratio
n
App
e
ndi
ces
Mai
n
tenan
c
e and
T
roub
le
sh
oo
tin
g
Com
m
is
si
on
in
g
10.23.3
Low pickoff voltage
Low pickoff voltage can be caused by several problems. See Table 10-6.
10.24
Checking sensor circuitry
Problems with sensor circuitry can cause several alarms, including sensor failure and a variety of
out-of-range conditions. Testing involves:
•
Inspecting the cable that connects the transmitter to the sensor
•
Measuring the resistances of the sensor's pin pairs and RTDs
•
Ensuring that the circuits are not shorted to each other or to the sensor case
Table 10-5
Drive gain problems, causes, and remedies
Cause
Possible remedy
Excessive slug flow
Entrained air
• Correct the process.
Cavitation or flashing
• Increase inlet or back pressure at the sensor.
• If a pump is located upstream from the sensor, increase the distance
between the pump and sensor.
Plugged flow tube
• Purge the flow tubes.
Mechanical binding of sensor tubes
• Ensure sensor tubes are free to vibrate. Possible problems include:
- Pipe stress. Check for pipe stress and eliminate if present.
- Lateral tube shift due to hammer effect. If this is a possibility,
contact Micro Motion.
- Warped tubes caused by overpressurization. If this is a possibility,
contact Micro Motion.
Incorrect sensor type configured
• Verify sensor type configuration, then verify sensor characterization.
Open drive or left pickoff sensor coil
• Contact Micro Motion.
Drive board or module failure, cracked flow tube,
or sensor imbalance
• Contact Micro Motion.
Table 10-6
Low pickoff voltage causes and remedies
Cause
Possible remedy
Slug flow
• See Section 10.17.
No tube vibration in sensor
• Check for plugging.
Moisture in the sensor electronics
• Eliminate the moisture in the sensor electronics.
Damaged sensor
• Ensure sensor is free to vibrate (no mechanical binding).
Possible problems include:
- Pipe stress. Check for pipe stress and eliminate if
present.
- Lateral tube shift due to hammer effect. If this is a
possibility, contact Micro Motion.
- Warped tubes caused by overpressurization. If this is a
possibility, contact Micro Motion.
• Test sensor circuitry. See Section 10.24.
• Contact Micro Motion.
Содержание 2200S
Страница 8: ...vi Micro Motion Model 2200S Transmitters ...
Страница 12: ...4 Micro Motion Model 2200S Transmitters ...
Страница 32: ...24 Micro Motion Model 2200S Transmitters ...
Страница 38: ...30 Micro Motion Model 2200S Transmitters ...
Страница 68: ...60 Micro Motion Model 2200S Transmitters ...
Страница 92: ...84 Micro Motion Model 2200S Transmitters ...
Страница 102: ...94 Micro Motion Model 2200S Transmitters ...
Страница 110: ...102 Micro Motion Model 2200S Transmitters ...
Страница 130: ...122 Micro Motion Model 2200S Transmitters ...
Страница 134: ...126 Micro Motion Model 2200S Transmitters ...
Страница 150: ...142 Micro Motion Model 2200S Transmitters ...
Страница 156: ...148 Micro Motion Model 2200S Transmitters ...
Страница 157: ......