56
Micro Motion
®
Model 2200S Transmitters
Configuring Operational Parameters
You can change the default severity level for some alarms. For example:
•
The default severity level for Alarm A020 (calibration factors unentered) is
Fault
, but you can
change the severity level to either
Informational
or
Ignore
.
•
The default severity level for Alarm A102 (drive over-range) is
Informational
, but you can
change the severity level to either
Ignore
or
Fault
.
For a list of all status alarms, default severity levels, and whether or not you can change the severity
level, see Table 5-5. (For more information on status alarms, including possible causes and
troubleshooting suggestions, see Table 10-3.)
Table 5-4
Status alarm severity levels
Severity level
Transmitter action
Fault
If this condition occurs, the alarm status flags are set, ALM_F flashes on the display, the
alarm is posted to the active alarm log, and the configured fault actions are performed.
Informational
If this condition occurs, the alarm status flags are set, ALM_I flashes on the display, and the
alarm is posted to the active alarm log, but fault actions are not performed.
Ignore
If this condition occurs, the alarm status flags are set, but no entry is added to the active
alarm log, the display does not show an alarm, and no fault actions are performed.
Table 5-5
Status alarms and severity levels
Alarm code
Communicator message
Default
severity
Configurable
Affected by
LMV Timeout
ProLink II message
A001
EEprom Checksum Error (Core Processor)
Fault
No
No
(E)EPROM Checksum Error (CP)
A002
RAM Test Error (Core Processor)
Fault
No
No
RAM Error (CP)
A003
Sensor Not Responding (No Tube Interrupt)
Fault
Yes
Yes
Sensor Failure
A004
Temperature sensor out of range
Fault
No
Yes
Temperature Sensor Failure
A005
Input Over-Range
Fault
Yes
Yes
Input Overrange
A006
Transmitter Not Characterized
Fault
Yes
No
Not Configured
A008
Density Outside Limits
Fault
Yes
Yes
Density Overrange
A009
Transmitter Initializing/Warming Up
Fault
Yes
No
Transmitter Initializing/Warming Up
A010
Calibration Failure
Fault
No
No
Calibration Failure
A011
Excess Calibration Correction, Zero too Low
Fault
Yes
No
Zero Too Low
A012
Excess Calibration Correction, Zero too High
Fault
Yes
No
Zero Too High
Содержание 2200S
Страница 8: ...vi Micro Motion Model 2200S Transmitters ...
Страница 12: ...4 Micro Motion Model 2200S Transmitters ...
Страница 32: ...24 Micro Motion Model 2200S Transmitters ...
Страница 38: ...30 Micro Motion Model 2200S Transmitters ...
Страница 68: ...60 Micro Motion Model 2200S Transmitters ...
Страница 92: ...84 Micro Motion Model 2200S Transmitters ...
Страница 102: ...94 Micro Motion Model 2200S Transmitters ...
Страница 110: ...102 Micro Motion Model 2200S Transmitters ...
Страница 130: ...122 Micro Motion Model 2200S Transmitters ...
Страница 134: ...126 Micro Motion Model 2200S Transmitters ...
Страница 150: ...142 Micro Motion Model 2200S Transmitters ...
Страница 156: ...148 Micro Motion Model 2200S Transmitters ...
Страница 157: ......