136
Micro Motion
®
Model 2200S Transmitters
Connecting with ProLink II Software
To make a HART/Bell 202 connection using ProLink II:
1. Attach the HART signal converter to the serial or USB port of your PC.
2. To connect to a HART multidrop network, connect the HART signal converter leads to any
point on the network (see Figure D-1). The HART/Bell 202 connection is polarity-insensitive.
To connect to the HART clips:
a.
Remove the transmitter housing cover (see Section C.3).
b.
Connect the HART signal converter leads to the HART clips on the face of the transmitter
(see Figure D-2). The HART/Bell 202 connection is polarity-insensitive. You can connect
either lead to either clip.
Note: The HART clips on the face of the transmitter are connected to the transmitter’s mA/HART
terminals. You may connect directly to the mA/HART terminals (terminals 1 and 2) if you have
removed the user interface module.
Figure D-1
HART/Bell 202 connections to network
VIA
TOR
DCS or
PLC
mA output wiring
Connection is polarity-insensitive
Содержание 2200S
Страница 8: ...vi Micro Motion Model 2200S Transmitters ...
Страница 12: ...4 Micro Motion Model 2200S Transmitters ...
Страница 32: ...24 Micro Motion Model 2200S Transmitters ...
Страница 38: ...30 Micro Motion Model 2200S Transmitters ...
Страница 68: ...60 Micro Motion Model 2200S Transmitters ...
Страница 92: ...84 Micro Motion Model 2200S Transmitters ...
Страница 102: ...94 Micro Motion Model 2200S Transmitters ...
Страница 110: ...102 Micro Motion Model 2200S Transmitters ...
Страница 130: ...122 Micro Motion Model 2200S Transmitters ...
Страница 134: ...126 Micro Motion Model 2200S Transmitters ...
Страница 150: ...142 Micro Motion Model 2200S Transmitters ...
Страница 156: ...148 Micro Motion Model 2200S Transmitters ...
Страница 157: ......