Page 66
Operation
capaNCDT 6100
Controller with option I (current output):
The values
- 4 mA (start of measuring range) and
- 20 mA (end of measuring range) are set for systems with current output.
5.4
Frequence Response
For slow dynamic measurements it is
profitable to reduce the cut-off frequen-
cy to 10 Hz. Effect:
- Noise decreases.
- Resolution increases.
Set the cut-off frequency to 10 Hz:
Open the controller.
Shift the switch S1 in the left positi-
on, see
Power/
output
Switch S1
2 kHz (Factory setting)
10 Hz
Fig. 9 Frequence Response
5.5
Synchronization in Multichannel Systems
Several measuring systems in the capaNCDT 6100 series can be operated simultaneously as multichannel
systems.
Synchronization of the measuring systems avoids interference between the sensors:
Plug SC30 synchronization cable (accessory) in the SYN OUT (synchronization output) socket in cont-
roller 1.
Plug the other end on the cable into the SYN IN (synchronization input) in controller 2.
Содержание 4350006.71
Страница 1: ...thicknessCONTROL MTS 8201 LLT O Frame Non contact thickness measurement Operating Instructions ...
Страница 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 95: ...Operating Instructions scanCONTROL 29xx ...
Страница 99: ...scanCONTROL 29xx Appendix A 1 Accessories 60 A 1 1 Recommended Accessories 60 A 1 2 Optional Accessories 61 ...
Страница 100: ......
Страница 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Страница 193: ......
Страница 194: ......
Страница 229: ......