Page 64
Operation
capaNCDT 6100
i
Disconnect the power supply before touching the sensor surface.
> Static discharge
> Danger of injury
The following assignment applies for a standardized setting value:
Trimmer position
Adjustment
Standardized value
Left stop
Minimum
0.00
Middle
Medium
5.00
Right stop
Maximum
10.00
5.3
Calibration with Metal Targets
Precondition - spec. resistance of the target < 1000 Ωcm.
Adjustment of the zero point and gain:
The controller is connected to the power supply and the display device according to the connection assign-
ment.
A linear characteristic is produced automatically due to the measuring principle and sensor construction.
The measuring device is set at the factory to 10 Volt FSO valid for sensor CS2 according to its measuring
range.
Sensors with other measuring ranges are recalibrated as follows:
Step 1
Open the cover of the controller.
Set the distance between the sensor surface and the target (use identical material for calibration and
operation) to start of measuring range.
Set the output voltage to 0 V with the ’zero’ potentiometer.
Содержание 4350006.71
Страница 1: ...thicknessCONTROL MTS 8201 LLT O Frame Non contact thickness measurement Operating Instructions ...
Страница 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 95: ...Operating Instructions scanCONTROL 29xx ...
Страница 99: ...scanCONTROL 29xx Appendix A 1 Accessories 60 A 1 1 Recommended Accessories 60 A 1 2 Optional Accessories 61 ...
Страница 100: ......
Страница 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Страница 193: ......
Страница 194: ......
Страница 229: ......