Page 51
Instruction Manual Software
thicknessCONTROL MTS 8201.LLT / 4350006.71
5.11.3 Method 3
Start measurement.
Change to menu
Extras > CGM Method 3.
Compares the first 40 measuring point of the first profile with the first 40 measuring
points of the second profile.
Fig. 61 Measurement of measuring device capability, Method 3
1
Result of measuring device capabilit
y
- valid after 80 points: %R&R < 20 %
2
Cursor shows the current position of the sensor unit.
3
Setpoint, upper and lower limit of production. Setpoint and threshold value are
used for calculation and diagrams.
4
This button starts and ends the test.
5
Shows the current state:
Test ongoing
Test completed, result not OK
Test completed, result OK
6
This diagram shows the current cross-profile.
Содержание 4350006.71
Страница 1: ...thicknessCONTROL MTS 8201 LLT O Frame Non contact thickness measurement Operating Instructions ...
Страница 12: ...Page 12 Laser Safety thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 62: ...Page 62 Instruction Manual Software thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 64: ...Page 64 Pneumatic Plan thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 70: ...Page 70 Declaration of Incorporation thicknessCONTROL MTS 8201 LLT 4350006 71 ...
Страница 95: ...Operating Instructions scanCONTROL 29xx ...
Страница 99: ...scanCONTROL 29xx Appendix A 1 Accessories 60 A 1 1 Recommended Accessories 60 A 1 2 Optional Accessories 61 ...
Страница 100: ......
Страница 135: ...Page 41 Operation of the Sensor with a PC scanCONTROL 29xx Fig 24 Predefined measuring fields ...
Страница 193: ......
Страница 194: ......
Страница 229: ......