77 - 77 021537/06/10
Command OS
<id>OS<para5> Scaling of wind velocity output (Output Scale)
Access: User mode
Description: This command is used to specify in which unit of measurement the wind velocity is
output in the serial telegram. Different units of measurement are available for this
purpose.
Parameter description:
Parameter
Unit of measurement Conversion factor related to m/s
0
m/s
1
1
km/h
1m/s => 3.6 km/h
2
miles/h
1m/s => 2.236936292 miles/h
3
knots
1m/s => 1.94253590 kn
Table 14: Conversion factors between different wind velocities
Range of values 0..3
Initial value: 0
Command PC
<id>PC<para5> Plausibility test (Plausibility check)
Access: User mode
Description: Switches the plausibility checks on/off. Every measured value is checked when
Plausibility is switched on. It is checked whether the measured value is plausible and
whether it fits into the sequence of measured values acquired. If this is the case, it is
released for further processing.
If this is not the case, the measured value is discarded. In the standard configuration
this has the following effects:
•
The heating is switched on: see
Command HT
The measured value acquisition interval is set to 0: see
Command MD
Range of values 0..7
Initial value: 7
Command PR
<id>PR<para5> Periodic receive time (Receive Time)
Access: Configuration mode
Description: This value sets the receive time period in the serial register in the analog section of the
ULTRASONIC. This value must not be changed under any circumstances. It is factory-
set.
Range of values 13..99
Initial value: 39
Unit: 100ns