64 - 64 021537/06/10
Command BH
<id>BH<para5> Selects the Bayern Hessen measuring instrument addresses (Bayern-Hessen
instrument address)
Access: User mode
Description: The ULTRASONIC contains several command interpreters. The THIES command
interpreter is active in standard mode. Command CI can be used to change the
command interpreter. If the Bayer - Hessen interpreter is selected, the measured
values for wind direction, wind velocity and virtual temperature are interrogated under
different measuring instrument addresses. The command BH specifies which
measuring instrument addresses are active. The following combinations are available
for selection:
Parameter description:
Parameter
value
Measuring
instrument address
of WV
Measuring
instrument address
of WD
Measuring
instrument address
of VT
0
11
1
22
1
400
410
420
2
202
201
211
3
11
1
21
4
81
82
83
5
1
2
3
Table 9: Measuring instrument addresses in Bayern Hessen command interpreter
See also Command CI, Bayern Hessen mode.
Command BL
<id>BL<para5> Returns the software version of the bootloader
Access: Query mode
Description: On query of the parameter the software version of the bootloader is returned..
For interpretation the output version must be divided by 100. An output value
00BL00300 represents the version V3.00.
Number ‚3’ means the main version, 00 is the build–label.
Command BP
<id>BP<para5> Determine the pretrigger in the burst mode.
Access User mode
Description: States the time (ms) of the data recording in the burst mode before the real trigger
event..
Value range: 0..65535
Initial value: 100ms