Service and Parts
|
6-7
Spare Analyzer Fuses
AMETEK recommends having on-site spare fuses used in the analyzer.
These fuses do not need to be changed out at regular intervals, but
are necessary in the event that a fuse has blown and a replacement is
required.
Analyzer Fuses,
120 VAC/240 VAC
120 V
240 V
Description
Part No.
Fuse Type
Part No.
Fuse Type
Analyzer
300-9524
1.6 A
300-4189
0.8 A
(TB2-1)
Over-Temp Power Supply
300-1519
63 mA
300-5097
32 mA
(TB2-2)
Oven Heater
300-4550
4 A
300-4548
2 A
(TB2-3)
Termination Solenoids
300-4189
0.8 A
300-5557
0.4 A
(TB2-4)
Calibration Solenoids
300-4189
0.8 A
300-4189
0.8 A
(TB2-5)
O2 Sensor
300-5465
3.15 A
300-5924
1.6 A
(TB2-6)
(*) 24 VDC Power Supply
300-4189
0.8 A
300-5557
0.4 A
(TB2-7)
(*) Note: Purgeable Deluxe (PD) (GP/Div 2 only) analyzers do not use the 24 V Power Supply (or 24 V fuse).
These fuses are located at
TB2
(see “AC Wiring” drawing in Chapter 3).
Sample and Vent Line (or SCU) Fuses
Part No.
Fuse Type
300-9291
6 A
300-9443
8 A
300-9292
10 A
300-9293
16 A
300-6312
20 A
Sample Line fuse (if used) is located at
TB3-30
on GP/Div 2 analyzers;
TB5-1
on Zone 1 analyzers.
Vent Line fuse (if used) is located at
TB3-50
on GP/Div 2 analyzers;
TB6-1
on Zone 1 analyzers.
See “Sample/Vent Line Wiring” drawings (GP/Div 2 Analyzers or Disconnect Enclosure, Zone 1 Analyzers) in Chapter 3, or
“Lower Cabinet Wiring” drawing in Appendix A.
The Sample and Vent Line fuses are typically not used; however, if
fuse terminals are retrofitted in the field for these temperature zone
circuits, the fuse required will depend on line length, voltage, and
power consumption. If the original factory-installed or retrofitted fuse
blows, remove it to determine the fuse type required.
NOTE
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...