Maintenance and Troubleshooting
|
5-5
Frequency Task
9–12 Months
Source Lamps
Replace the source lamps.
The lamps may require replacement sooner if analyzer alarms are
set (“
w Lamp Control
” or “
w PMT Signals
”). See also “Source Lamp
Replacement” and “Troubleshooting and Diagnostics” in this chapter.
Every Year
Measuring Cell
Replace the o-rings and Cell Windows and clean the interior surfaces to
remove any contaminants.
This schedule is a minimum requirement. If other conditions are found to
be present (see “Measuring Cell Maintenance”) more frequent cleaning
may be required.
Every 2 Years
Chopper Assembly, Optical Bench
Replace the Chopper Motor Drive Belt and Chopper Wheel Bearings.
If there are recurring analyzer alarms that suggest problems with the
Chopper Wheel or other related parts in the Chopper Assembly, replace
these parts sooner (that is, “
w Lamp Control
”, “
w PMT Signals
”,
“
f Wheel Speed
”, or “
w ZDrift/SpnErr
”).
See “Chopper Assembly Maintenance” in this chapter.
Other
Heated Acid Gas (HAG) Sample Probe (Optional)
Replace the o-rings and filters.
Refer to the HAG Sample Probe
Installation and Maintenance Guide
for
recommended preventive maintenance requirements.
Other
Sample/Vent Isolation Valves
The isolation valves in the Sample Probe (if used, or SCU) and Vent
Valve (if used) assemblies typically do not require regular maintenance.
However, if it is suspected that leaks are originating from either of these
valves, change out the appropriate valve assembly immediately. Contact
AMETEK for more information.
Other
Flamepath Gap
During each analyzer maintenance, use a feeler gauge to check the
flamepath gap of the following locations. The gap must not exceed
the listed gap for each location;
if the gap exceeds this value,
contact AMETEK for advice
. See also “Examining and Caring For the
Flamepaths” in this chapter.
• Disconnect Enclosure flange (enclosure door and housing joining
surfaces).
Maximum Flamepath Gap:
0.15 mm
• Heater Plate Assembly flange (upper/lower Heater Plate joining
surfaces, inside the Analyzer Oven).
Maximum Flamepath Gap:
0.1 mm
For other flamepaths not mentioned here, contact AMETEK.
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...