4-16 | Model 920 Multi-Gas Analyzer
F1 5
PmtBal
** A secondary PMT control signal (V) which adjusts the PMT gain to equalize the PMT sig
-
nals when the filters with the lowest and highest transmittances are in the light path. Normal
values after a successful Auto-Setup are from 4.0–7.0 VDC.
F1 6
SigMax
** The signal from the PMT with the highest gain is set to SigMax when each filter is in
the light path while the Auto-Setup is in progress. SigMax should be 8.0–9.0 V (check
EEPROM Data Sheets
for actual value).
F1 7
Kf
z
The units conversion constant for calculation of component mass flow (Mass/Energy),
where ‘
z
’ = 1..7.
1 = Kf1
5 = Kf5
2 = Kf2
6 = Kf6
3 = Kf3
7 = Kf7
4 = Kf4
F1 8
Ke
z
The units conversion constant for calculation of component mass flow (Mass/Amount),
where ‘
z
’ = 1..7.
1 = Ke1
5 = Ke5
2 = Ke2
6 = Ke6
3 = Ke3
7 = Ke7
4 = Ke4
F1 9
Units
Units of measurement used for displaying concentrations, temperatures, and pressures.
The value is the sum of five bits:
Bit 0
Value
Definition
1
0 = Metric Units
1 = Imperial Units
Bits 2,1
Value
Definition
2,4*
0 = Normal concentration result (PPM).
2 = Normal concentration result (%).
4 = Finer (Low) concentration result (PPM).
6 = Coarse (High) concentration result (%).
*Sum of Bits 1 and 2 decimal value
Bit 3
Value
Definition
8
0 = Display 4 items.
8 = Display 3 items.
Bit 4
Value
Definition
16
0 = Remote Auto-CAL
16 = Remote Backpurge/Zero Flow
For example, if the analyzer is configured for Metric units, Normal concentration result in
PPM, Display 3 items, Remote Auto-CAL, the result would be:
Bit 0 + Bits 2,1 + Bit 3 + Bit 4 [0 + 0 + 8 + 0 = 8].
The unit for input and output parameters (Metric or Imperial) is set at the factory, and
should not be changed (i.e., do not change from a Metric unit to an Imperial unit). If a
change is required, consult with AMETEK.
You can, however, change the concentration units freely back and forth between PPM and
mole percent (e.g., from Units = 0 to Units = 2, or from Units = 1 to Units = 3, or vice versa).
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...