4-34 | Model 920 Multi-Gas Analyzer
CAL Mode – F2 Commands
Keystroke Term
Definition
F2 •
—
—
F2 –
—
—
F2 0
Man/Zero?
Allows you to manually Zero the analyzer. The analyzer Zero is automatically adjusted
based on the average reading during the
IntTime
duration after a
Yes
response only
if
AdjDisable
is set to ‘
0
’. The Zero gas must be introduced manually into the analyzer
sample system. Refer to “Manually Zeroing the Analyzer” in Chapter 3.
F2 y
Man/Spany? Allows you to manually Span the analyzer for each component, where ‘y’ = Manual Span
F2 1..F2 6
. The analyzer calibration is automatically adjusted based on the average read-
ings during the
IntTime
duration after a
Yes
response only if
AdjDisable
is set to ‘
0
’. The
Calibration (Span) gas must be introduced manually into the analyzer sample system.
Refer to “Manual Span” in this chapter for details about how to manually Span the analyzer.
F2 1 = Man/Span1?
F2 4 = Man/Span4?
F2 2 = Man/Span2?
F2 5 = Man/Span5?
F2 3 = Man/Span3?
F2 6 = Man/Span6?
F2 7
—
—
F2 8
Man/SpanO2? Allows you to manually Span the O
2
Sensor (if used –
General Purpose applications/
analyzers only
) at the analyzer. The sensor Calibration is automatically adjusted based on
the average reading during the
IntTime
duration after a
Yes
response only if
AdjDisable
is
set to ‘
0
’. Man/Span must be used to calibrate the sensor if Nitrogen is used for the Span
gas. The Calibration (Span) gas must be introduced manually into the analyzer sample sys-
tem. For systems with the O
2
Sensor, refer to the
O
2
Concentration Measurement Manual
Supplement
for details about how to manually Span the O
2
Sensor.
F2 9
Man/ZeroO2? Allows you to manually Zero (air point calibration) the O
2
Sensor (if used –
General
Purpose applications/analyzers only
) at the analyzer. The sensor Zero is automatically
adjusted based on the average reading during the
IntTime
duration after a
Yes
response
only if
AdjDisable
is set to ‘
0
’. Man/Zero must be used to calibrate the sensor if Nitrogen is
used for the Zero gas. The Zero gas must be introduced manually into the analyzer sample
system. For systems with the O
2
Sensor, refer to the
O
2
Concentration Measurement
Manual Supplement
for details about how to manually Zero the O
2
Sensor.
Содержание 920
Страница 39: ...Specifications 2 11 ...
Страница 40: ...2 12 Model 920 Multi Gas Analyzer ...
Страница 41: ...Specifications 2 13 2019 02 05 ...
Страница 42: ...2 14 Model 920 Multi Gas Analyzer ...
Страница 43: ...Specifications 2 15 ...
Страница 44: ...2 16 Model 920 Multi Gas Analyzer ...
Страница 45: ...Specifications 2 17 ...
Страница 47: ...Specifications 2 19 Heater Plate ATEX and IECEx Certificates and Marking ...
Страница 48: ...2 20 Model 920 Multi Gas Analyzer ...
Страница 49: ...Specifications 2 21 ...
Страница 50: ...2 22 Model 920 Multi Gas Analyzer ...
Страница 51: ...Specifications 2 23 ...
Страница 52: ...2 24 Model 920 Multi Gas Analyzer ...
Страница 53: ...Specifications 2 25 ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 54: ...2 26 Model 920 Multi Gas Analyzer Disconnect Enclosure Type 200 ATEX and IECEx Certificates and Marking ...
Страница 55: ...Specifications 2 27 ...
Страница 56: ...2 28 Model 920 Multi Gas Analyzer ...
Страница 57: ...Specifications 2 29 ...
Страница 58: ...2 30 Model 920 Multi Gas Analyzer ...
Страница 59: ...Specifications 2 31 ...
Страница 60: ...2 32 Model 920 Multi Gas Analyzer ATEX and IECEx certified Model 920 Analyzers are marked with this label ...
Страница 71: ...Installation and Start Up 3 11 Figure 3 4 Optical Bench Measuring Cell assembly ...
Страница 99: ...Installation and Start Up 3 39 Figure 3 11 Over Temp alarm reset switch SW300 Termination board ...
Страница 108: ...3 48 Model 920 Multi Gas Analyzer This page intentionally left blank ...
Страница 173: ...Controller User Interface 4 65 Figure 4 8 Customer signal connections ...
Страница 187: ...Maintenance and Troubleshooting 5 11 Figure 5 1 Measuring Cell assembly ...
Страница 211: ...Maintenance and Troubleshooting 5 35 Figure 5 5 Optical Bench Chopper Assembly Maintenance Diagram ...
Страница 226: ...5 50 Model 920 Multi Gas Analyzer Figure 5 8 Flameproof Heater Seal Assembly layout WX 14324 1A ...
Страница 227: ...Maintenance and Troubleshooting 5 51 Figure 5 9 Flameproof Heater Seal internal component layout 100 1622 1A ...
Страница 228: ...5 52 Model 920 Multi Gas Analyzer Figure 5 10 Oven Heater and Temperature Sensor wiring WX 14161 ...
Страница 243: ...Maintenance and Troubleshooting 5 67 Figure 5 13 Over temp alarm reset switch SW300 Termination board ...
Страница 262: ...A 2 Model 920 Multi Gas Analyzer Ribbon Cable Interconnect WX 102836 ...
Страница 263: ...Appendix A Drawings A 3 GP Lower Enclosure to Electronics Wiring CE Analyzers WX 102810 ...
Страница 264: ...A 4 Model 920 Multi Gas Analyzer Heater and Sensor Wiring GP Div 2 Analyzers WX 102851 ...
Страница 265: ...Appendix A Drawings A 5 Heater and Sensor Wiring CE Zone 1 Analyzers WX 102852 ...
Страница 266: ...A 6 Model 920 Multi Gas Analyzer Lower Cabinet Wiring CE GP Analyzers 120V 100 1341 3 ...
Страница 267: ...Appendix A Drawings A 7 Lower Cabinet Wiring CE GP Analyzers 240V 100 1342 3 ...
Страница 268: ...A 8 Model 920 Multi Gas Analyzer Signal Wiring PD GP Div 2 CE Zone 1 Analyzers WX 102815 ...
Страница 270: ...A 10 Model 920 Multi Gas Analyzer 15V and 5V Power Supply DC Wiring GP Div 2 CE Zone 1 Analyzers WX 102811 ...
Страница 271: ...Appendix A Drawings A 11 24V Power Supply DC Wiring CE Zone 1 Analyzers WX 102812 ...
Страница 272: ...A 12 Model 920 Multi Gas Analyzer RS 232 Communications Cable Wiring 300 9480 ...
Страница 273: ...Appendix A Drawings A 13 RS 232 RS 485 Module Wiring CE Zone 1 GP Div 2 Analyzers 100 2185 ...
Страница 274: ...A 14 Model 920 Multi Gas Analyzer Microcontroller Board 100 0117 ...
Страница 275: ...Appendix A Drawings A 15 Host Controller Board Display Interface 100 0138 ...
Страница 276: ...A 16 Model 920 Multi Gas Analyzer Model 9xx Series Analyzer Type 200 Disconnect Enclosure Details ...
Страница 278: ...S 2 Model 920 Multi Gas Analyzer This page intentionally left blank ...