Appendix C: Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding
electromagnetic compatibility. Ensure that the
Arctic Sun
®
Temperature
Management System is installed and used according to the electromagnetic
compatibility information provided. The following are guidance and
manufacturer’s declarations regarding electromagnetic compatibility for
the
Arctic Sun
®
Temperature Management System.
• The use of accessories or cables other than those specified or sold by
Medivance (shown below) is not recommended. Use of unapproved
accessories or cables may result in increased emissions or in decreased
immunity of the
Arctic Sun
®
Temperature Management System.
• If the
Arctic Sun
®
Temperature Management System is used directly
adjacent to or stacked with other equipment, the user should periodically
observe the
Arctic Sun
®
Temperature Management System device to
verify it operates normally in that environment.
• Portable and mobile RF communications equipment can affect Medical
Electrical Equipment.
Accessories and cables approved by Medivance for use with
the
Arctic Sun
®
Temperature Management System
Part #
Temperature In-Cable - Nellcor
735-02
Temperature In-Cable - Bard
®
735-03
Temperature In-Cable - Rusch
735-04
Temperature In-Cable - GE
735-05
Temperature In-Cable - Philips
735-06
Temperature Out-Cable - Nellcor
735-52
Temperature Out-Cable - Bard
®
735-53
Temperature Out-Cable - Rusch
735-54
Temperature Out-Cable - GE
735-55
Temperature Out-Cable - Philips
735-56
Power Cord, US, Canada, Mexico
733-00
Power Cord, Continental Europe
733-01
Power Cord, UK, Ireland
733-02
Power Cord, Australia, New Zealand
733-03
Power Cord, Mainland China
733-04
Power Cord, Brazil
733-05
Power Cord, Switzerland
733-07
Power Cord, South Africa
733-08
Transmission Interface Module (TIM) Kit
760-00
Transmission Interface Module (TIM)
761-00
RS232 Cord
762-00
1.1 EN/IEC 60601-1-2 Table 1
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The
Arctic Sun
®
Temperature Management System is intended for use in the
electromagnetic environment specified below. The customer or the end user of the
Arctic Sun
®
Temperature Management System should assure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
Arctic Sun
®
Temperature
Management System uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
1.1 EN/IEC 60601-1-2 Table 1 (cont.)
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Class A
The
Arctic Sun
®
Temperature
Management System unit is suitable
for use in all establishments other
than domestic, establishments and
those directly connected to the public
low-voltage power supply network
that supplies buildings for domestic
purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
Flicker emissions IEC
61000-3-3
Complies
1.2 EN/IEC 60601-1-2 Table 2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
Arctic Sun
®
Temperature Management System unit is intended for use in the
electromagnetic environment specified below. The customer or the end user of the
Arctic Sun
®
Temperature Management System unit should assure it is used only in
such an environment.
Immunity Test
IEC60601 test
level
Compliance
Level
Intended
Electromagnetic
Environment
Electromagnetic
discharge (ESD)
IEC 61000-4-2
±
6kV contact
±
8kV air
±
6kV contact
±
8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2kV for power
supply lines
±
1kV for input/
output lines
±
2kV for power
supply lines
±
1kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±
1kV differential
mode (line-line)
±
2kV common
mode (line-earth)
±
1kV differential
mode (line-line)
±
2kV common
mode (line-earth)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT (>95%
dip in UT) for
0.5 cycle
40% UT (60%
dip in UT) for
5 cycles
70% UT (30%
dip in UT) for
25 cycles
<5% UT (>95%
dip in UT) for
5 seconds
<5% UT (>95%
dip in UT) for
0.5 cycle
40% UT (60%
dip in UT) for
5 cycles
70% UT (30%
dip in UT) for
25 cycles
<5% UT (>95%
dip in UT) for
5 seconds
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
Arctic Sun
®
Temperature
Management System
unit requires continued
operation during power
mains interruptions, it is
recommended that the
Arctic Sun
®
Temperature
Management System
unit be powered from
an uninterruptible
power supply with
sufficient capacity to
run the unit for the
maximum required time
of interruption.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Содержание ARCTIC SUN Simply Advanced 5000
Страница 1: ...TEMPERATURE MANAGEMENT SYSTEM SimplyAdvanced M O D E L 5 0 0 0 O P E R A T O R S M A N U A L ...
Страница 13: ...13 E N G L I S H ...
Страница 25: ...25 F R A N Ç A I S F R E N C H ...
Страница 121: ...121 D A N S K D A N I S H ...
Страница 133: ...133 S V E N S K A S W E D I S H ...
Страница 145: ...145 S U O M I F I N N I S H ...
Страница 157: ...157 N O R S K N O R W E G I A N 157 ...
Страница 193: ...193 193 Č E S K Y C z e c h ...
Страница 205: ...205 T Ü R K Ç E T u r k i s h ...
Страница 217: ...217 R O M Â N Ă R o m a n i a n ...
Страница 229: ...229 S L O V E N Č I N A S l o v a k i a ...
Страница 273: ...273 한 국 어 K O R E A N ...