4-21
The Deployment parameters displays next. These settings will be applied globally to all events in
the deployment.
Figure 4.6-11: Header and Optional Tilt Settings
•
Headers
: Options A, B, and C of the Event Parameters allow the user to define three
headers to describe the deployment.
•
Sensor sampling interval
: If the Sediment Trap is configured with the optional tilt sensor,
the sensor sampling interval can be changed from the default of 1 minute.
10.
Type
v
to proceed with the deployment.
Proceeding with the Deployment
After the status line (and battery warnings if any), a final prompt displays.
Figure 4.6-12: Status Line
With firmware v3.12 and higher, electronics hardware changes made the
backup battery unnecessary and it was elminated. If deploying a Sediment Trap
that still has a backup battery, install a fresh battery before every deployment.
The pre-deployment program confirms backup batteries are installed, and if not
detected, the firmware will not accept the ‘Y’ entry to start the deployment.
Performing 6 second Backup Battery test...
System status:
Date Time Battery Int. Temp Rotator
10/03/16 06:54:18 21.0 Vb 22.8
∞
C aligned
Caution: Deployment will overwrite the EEPROM data backup cache.
Proceed with the deployment [N] y
Setting for optional
compass/tilt
Proceed with the deployment
Header A| System Test
B| ML 12345-01
C| MWB
D| Sensor sampling interval [minutes]: 1
V| Verify and proceed.
Selection [ ] ? v
Содержание Mark 78H-21
Страница 1: ...Sediment Trap User Manual...
Страница 6: ...TOC 4...
Страница 28: ...1 18 Notes...
Страница 44: ...3 8 Notes...
Страница 74: ...5 8 Notes...
Страница 104: ...A 4 Notes...
Страница 128: ...A 4 Notes...
Страница 151: ...Appendix D 3 Test Cable Wiring Diagram Figure D 3 Test Cable Wiring Diagram...
Страница 160: ...Appendix E 8 Notes...
Страница 168: ...Appendix E 8 Notes...