Appendix C-2
ST - Current Status
The ‘ST’ command displays Event #, confirms rotator alignment, shows electronics temperature
in °C, battery voltage, previous sample date/time, current date/time, next sample date/time
(shown in Figure C-8).
ML12294-02 > ST
2 Y 23.8 25 05/15/2008 10:00:00 05/15/2008 10:00:38 05/15/2008 10:05:00
Figure C-2: ‘ST’ Current Status
RB - Rotate Bottle
‘RB’ Rotate Bottle commands the firmware to rotate to the next bottle. There is a 5 minute delay
between the command and the rotation to allow the user to make corrections. As shown in Figure
C-9, no other events are changed. A status line is returned to show the change.
05/15/2008 10:03:06 Sleeping . . .
ML12294-02 > RB
2 Y 23.8 25 05/25/2008 12:00:00 05/25/2008 10:03:16 05/25/2008 10:08:16
05/15/2008 10:03:18 Sleeping . . .
ML12294-02 > VE ALL
1 of 22 = 05/23/2008 12:00:00
2 of 22 = 05/25/2008 10:08:16
3 of 22 = 05/27/2008 12:00:00
4 of 22 = 05/29/2008 12:00:00
5 of 22 = 05/31/2008 12:00:00
6 of 22 = 06/02/2008 12:00:00
Figure C-3: ‘RB’ Rotate Bottle
Even
t
Alignme
nt
Voltage
Elec
Temp
Previous Sample
Date/Time
Current
Date/Time
Next Sample
Date/Time