4-22
11.
Type
Y
to commit to the deployment (or
N
to cancel and return to the Main Menu).
Prior to submerging the sampler, disconnect the communications cable from the
controller housing and reconnect the dummy plug to the communications
connector on the controller housing.
12.
The Sediment Trap enters a low power sleep mode until the time of the first
scheduled event. The Sediment Trap automatically wakes up and begins sampling
at the scheduled time of the first event.
>>> Remove communication cable and attach dummy plug. <<<
>>> System is ready to deploy. <<<
03/06/15 10:38:46 Waiting for event 01 of 05 @ 03/06/15 11:30:00
03/06/15 10:38:46 Suspended until 03/06/15 11:30:00 ....
Figure 4.6-13: Deploy System
13.
For a bench test, leave the communications cable attached to the Sediment Trap
electronics to observe the operation of the instrument in the McLaneTerm window.
14.
For a deployment, proceed to Chapter 5 and follow the Deployment Preparation
steps to disconnect the communications cable, seal the end cap, and attach the
dummy plug.