Appendix B-5
Configuration Menu –Atmospheric Pressure Offset
The menu option <P> displays for RBR loggers that will record pressure. Selecting this option
allows you to zero the atmospheric pressure either automatically or manually. This sets the
current ambient pressure on the RBR logger to zero scale at the surface so that the logger reading
reflects the pressure from below the surface.
Figure B-5: Option <P> Atmospheric Pressure Calculation
This is typically only necessary if deploying in a location that is not near sea
level (for example, a high mountain lake).
Figure B-6: Automatic Atmospheric Pressure Offset
Typing
a
calibrates the pressure automatically. Automatic calibration takes the average of several
pressure readings for the atmospheric pressure at the sampler’s current location.
________________________________________________________________
Configuration: PST-21_CT_RBR-TD CF2 V3_11 of Aug 11 2015
_________________________________
Configuration Menu
_________________________________
Wed Aug 12 13:31:07 2015
<A> Number Of Cups [21]
<B> Compass Tilt [Yes]
<C> Integral Temperature Sensor [No]
<D> Integral Pressure Sensor [No]
<E> RBR duo TD [Yes]
<P> Atmospheric Pressure Zero Offset
<X> Save & Exit
<^C> Cancel & Exit
Selection [ ] ? p
Retreiving atmospheric pressure value
Current atmospheric pressure value is set at 10.6000
Change the atmospheric pressure value? [N] ? y
Manually enter value [M], or determine it automatically [A]? [] ? A
Determining the current atmospheric pressure . . . . .
Setting the atmospheric pressure value.
Atmospheric Pressure = 10.1647
Содержание Mark 78H-21
Страница 1: ...Sediment Trap User Manual...
Страница 6: ...TOC 4...
Страница 28: ...1 18 Notes...
Страница 44: ...3 8 Notes...
Страница 74: ...5 8 Notes...
Страница 104: ...A 4 Notes...
Страница 128: ...A 4 Notes...
Страница 151: ...Appendix D 3 Test Cable Wiring Diagram Figure D 3 Test Cable Wiring Diagram...
Страница 160: ...Appendix E 8 Notes...
Страница 168: ...Appendix E 8 Notes...