2-7
Backup batteries are not installed with firmware v3.12 and higher. Electronics
hardware changes made the backup battery unnecessary and it was elminated. If
deploying a Sediment Trap that still has a backup battery, install a fresh battery
before every deployment.
Establishing Communication
1.
Find the communication cable provided in the toolkit.
2.
Connect the communications cable to the computer serial communication port
before connecting to the communications connector on the controller end cap.
Figure 2-8: Connecting the COM Cable
3.
Align the bulkhead connection pins, and push the bulkhead in place on the
controller end cap.
4.
On the computer, open a properly configured McLaneTerm window.
5.
Connect the DB9 end of the communication cable to the computer communication
port configured in McLaneTerm.
6.
Enter [CTRL]-[C] in McLaneTerm. If the computer setup procedures were
completed correctly, the Main Menu, a system clock confirmation message, or a
message that indicates the system is Suspended will print to the screen.