Manual
maMoS
100/200/300/400
ATTENTION!
It is forbidden to use a mixture of gases to execute calibration pro-
cess. If the analyser is equipped with the sensor which detects any
gas being a component of the mixture the calibration will never be
performed properly.
4. Press "
Hold readouts"
button, and then "
Calibrate".
In the left corner of the
status bar you can observe the progress of calibration data being saved.
5. After several seconds a calibration curve will appear. If the chart is illegible
close it and reopen it using "
Chart"
button.
6. If a new curve is incorrect or if the sensor starts working improperly you can
restore the factory curve using "
Factory"
button.
7.3.11.2. Electrochemical sensor calibration
If the “
CPU: Mamos III
” or
“SensElch”
slot is chosen from the
“1-point calibration”
menu
then the following window appears:
Picture 106.
1-point calibration window for the electrochemical sensors
Window shows all sensors available in the chosen channel. To calibrate a sensor
follow the instructions:
1. Interrupt running measurements by pressing "
Stop"
button. Current results
will no longer be visible and buttons referring to installed sensors will become
active.
2. Apply the fresh air to the gas inlet or switch the sensor into a
“VENTILA-
TION”
phase (using cycle's reset function (see chapter 7.3.8. ). Ventilation
should last for about 15 minutes.