7.3.11.1. IR sensor calibration
After the sensor which is to be calibrated has been chosen from the „1-point calib-
ration” menu, the window shown below will appear.
Picture 104.
The window containing the IR sensor readings.
To execute calibration:
1. Apply a neutral (for the specified sensor) gas to the device and wait for the
values in fields "
Ch1"
, "
Norm."
and "
Ch2"
to stabilize. Press
“Hold readouts”
button
.
Picture 105.
The readings if the zeroing gas is supplied.
2. Enter the concentration of the neutral gas into
“Zeroing gas”
field, and then
press
“Make zeroing”
button.
3. After zeroing, apply the reference gas and enter it's concentration into
“Calib-
ration Gas"
field. Then, press
“Read"
and wait for the values in fields "
Ch1”
,
"
Norm."
and "
Ch2"
to stabilize.
93