Manual
maMoS
100/200/300/400
can be checked. After the button has been pressed the state of the relay output will be re-
placed with the opposite one for 8 seconds. After that time the actions of the outputs will
be performed according to the settings.
Picture 101.
The settings of the relay in case the IN1 inlet is chosen for the source.
The choice of the source signal controlling the output can be made with the help of
the list located below the picture. The choice can be made between the analogue outputs,
the digital outputs (labelled as
"Follows IN1"
and
"Follows IN2"), the "Follows Measure-
ment"
, the
“Flow control”
position and the
"No action"
position. If the
"No action"
option is
chosen, the output is switched off (it remains inactive). If the
“Flow control”
option is selec-
ted then the relay reacts to the flow value in the gas channel. If the flow drops below the
critical level the relay opens. Choosing the
"Follows Measurement"
switches the output on
when the monitor proceeds to the
"MEASUREMENTS"
stage. The output is switched off
during all other stages. If the user chooses the
"Follows IN1"
or
"Follows IN2"
the appropri-
ate relay is activated when the input (IN1 or IN2) is in active stage. As the state on the in-
90