Manual
maMoS
100/200/300/400
Due to the
FAT-16
file system restrictions the main catalogue can consist of up to
512 files. It the number is exceeded an error occurrence will be reported (the LED will be
flashing – see point 5.4.1. ).
The
MMC
module of the
maMoS
monitor works in several stages – each of them
signalised with an appropriate flash code of the LED (see point 5.4.1. ).
No card
– no card is in slot (the LED is constantly off).
Initialisation
– performed when the card is inserted into the slot or immedi-
ately after the recording starts. This stage lasts for about 5 s during which the
card parameters are read and (eventually) a new file used for recording the
data on the card is created (the LED flashes ten times per second).
Off
– module proceeds to this stage if the recording is stopped (the LED
flashes once every three seconds).
Recording
– during this stage the measurement results are periodically re-
corded on the card (the LED is constantly on). The recording frequency can
be set with the use of the
MaMoSII.exe
software.
Stopping
– during this stage the
MMC
module closes the current file (the
LED flashes ten times per second) and, according to the situation:
−
proceeds to the
Initialisation
stage – if the maximum file size has been
exceeded (10 000 records),
−
proceeds to the
Off
stage – if the recording process has been stopped
with the help of the computer program,
−
proceeds to the
Stopped
stage – if the
Hold recording
button has been
pressed.
Stopped
– stays on hold for 10 s in order to make it possible for the user to
remove the card (the LED is off). If the card is not removed the recording will
be automatically restarted.
Error
– the module proceeds to this stage if it detects any errors in the card's
working (the LED flashes once per second). In this case storing and reading
the data from the card is impossible.
Before the card is removed from the slot the user must press the
Hold recording
button for at least two seconds and release it. The file with the results will then be closed
and the recording will stop (the LED will be flashing quickly and then it will turn off). The
card can then be safely removed. If the card is not taken out of the slot the
Initialisation
and
Recording
will start after 10 seconds. During the
Recording
stage (when the LED