Manual
maMoS
100/200/300/400
7.3.3.3. Gas pump
Picture 76.
Analyser's settings – Gas pump
In the "
Pump capacity"
section the gas pump
capacity
can be set. The
capacity
can be set from 100 to 255 (100 is the minimum and 255 – the maximum pump speed (ef-
ficiency)).
Section
“Flow control”
is active only when the device is equipped with the differen-
tial pressure sensor that allows to control the flow in gas channel. This function allows to
control the capacity of dryer's filter and permeability of the hose supplying the gas
samples.
"Enable control"
option turns on/off the flow control. In window "
Minimum flow level
[l/h]"
user can enter the minimum flow value below which the analyser will signalise that
the flow is to low.
If the flow control is turned on and the analyser is in one of the cycles which re-
quires the use of pump (“VENTILATION”, “MEASUREMENTS” or “VENTILATION BE-
FORE STAND-BY”), and this cycle lasts for more than 30s then the analyser measures the
68