Manual
maMoS
100/200/300/400
11. To check whether the programming process has been successful the user
should:
−
switch the monitor off
−
remove the jumper
−
switch the monitor on and confirm whether the firmware version is correct
(appears on the display
#1
several seconds after the device has been
switched on
)
. If the firmware version is correct the monitor can be switched
off and the casing of the device should be closed.
Picture 52.
Information of the completion of the updating process.
5.6. Flow velocity measurement
MaMoS
analyser can optionally measure gas flow velocity (for example in chim-
neys). If this option is available then the analyser is equipped with the pressure measure -
ment inlet and outlet. They are installed at the bottom part of the housing next to fresh air
inlet and gas outlet.
The measurement of the flow velocity is made via measurement of total and static
pressure using Pitot tube. To perform the measurement correctly connect the Pitot tube as
follows:
The outlet of the static pressure connect to the inlet signed with
symbol
The inlet of the total pressure connect to the inlet signed with
symbol
49