22  Regulatory Compliance  

 

Product Safety: 

 

IEC 61010-1:2001

 

Safety Requirements for electrical Equipment for 
Measurement, Control and Laboratory Use Part 1.  

 
EN 61010-1:2001

 

 

 

Safety Requirements for   
Electrical Equipment for Measurement, Control and 
Laboratory Use. 

 
CAN/CSA C22.2 No. 61010-1-04

/

 

Safety Requirements for  

UL 61010-1: 2004

 

 

 

Electrical Equipment for  

 

 

 

 

 

Measurement, Control and  
Laboratory Use Part 1; General Requirements. 

 

Electrical Equipment for Laboratory Use Part 1: 
General Requirements 

 

IEC Equipment Class:

   

 

 

Installation Category:

    

 

II 

 

Pollution Degree:

      

 

 

Electromagnetic Compatibility: 
EN 61326-1: 2006

 

Electromagnetic Compatibility Immunity Testing-
Measurement, Control and Laboratory Equipment  

 

EN55011:2007/CISPR11:2004  
Class B, Group 1  
Emissions Testing:   

 

Industrial, scientific and Medical Equipment 

 
CE Marking: 
Council Directive 2006/95/EC

 

Low Voltage Directive 

Council Directive 2004/108/EC

 

EMC Directive 

Council Directive 2002/96/EC

 

WEEE Directive 

Council Directive 2006/95/EC

 

RoHS Directive 

 

Содержание OmniCure S2000

Страница 1: ...OmniCure S2000 USER S GUIDE Lumen Dynamics Group Inc 2260 Argentia Road Mississauga ON L5N 6H7 Canada 1 905 821 2600 www LDGI com Printed in Canada 035 00362R Rev 2 ...

Страница 2: ...tion in this manual is accurate however information in this manual is subject to change without notice and does not represent a commitment on the part of the authors Trademarks OmniCure and Intelli Lamp are trademarks of Lumen Dynamics Group Inc All other product names are trademarks of their respective owners S2000 Control Panel Software Minimum Computer Specifications 300 MHz processor Pentium o...

Страница 3: ...scriptions 20 11 3 Radiometer RS 232 Phono Connector Signal Descriptions 29 11 4 Audio Style Foot Pedal Connector Signal Descriptions 29 12 Viewing the Accumulated Lamp Hours 29 13 Messages Indicators 30 14 LED Indicators 33 15 Clearing Audible Alarm 34 16 Remote Automated Control Requirements 35 16 2 Com Port Configuration 35 16 3 Message Format and Protocol 35 16 4 Command Descriptions 36 16 5 C...

Страница 4: ...e Cleaning 58 20 Troubleshooting 60 21 Technical Specifications 64 21 1 Lamp Module 64 21 2 Light Guide 66 21 3 I O Ports including RS 232 67 21 4 USB Interface 67 21 5 Environmental Conditions 67 21 6 Noise and the OmniCure SERIES 2000 68 22 Regulatory Compliance 70 22 1 FCC Part 15 Subpart B Class B Unintentional Radiators 71 22 2 China RoHS 72 22 3 WEEE Directive 2002 96 EU 73 22 4 Mechanical S...

Страница 5: ...it Configuration 21 Figure 8 Rear Panel Sample Connection 22 List of Tables Table 1 P1 Pin Out 23 Table 2 P2 Pin Out 24 Table 3 P3 Pin Out 28 Table 4 RS 232 Phono Connector Pin Out 29 Table 5 Foot Pedal Pin Out 29 Table 6 LED Status 33 Table 7 Unit Status 44 Table 8 Lamp Part Number 57 Table 9 Replacement Filter Part Numbers 58 Table 10 Message References 60 Table 11 Front Panel LED Descriptions 6...

Страница 6: ... This extended lifetime is made possible by the proprietary technology incorporated into the design of the system The lamp is mounted in an elliptical reflector with a proprietary coating to provide excellent spectral and power output Control for the OmniCure SERIES 2000 is provided by a new rotary shutter iris mechanism The iris is adjustable in 1 increments to provide very precise amounts of lig...

Страница 7: ...ch 5 Lamp Housing Access Tool fastened beneath the system housing 6 Grounded POWER Cord 7 RS 232 Cable 9 Pin 8 CD with User Guide and Interface Software If your packaged unit is missing any of the above components call Lumen Dynamics at 905 821 2600 or 1 800 668 8752 Any additional optional items purchased to customize the unit will also be present ...

Страница 8: ...S LED LIGHT GUIDE STATUS LED LIGHT GUIDE PORT POWER ON OFF SWITCH START STOP BUTTON SHUTTER ACTIVATION CLOSED LOOP FEEDBACK LED RADIOMETER CALIBRATION PORT RS 232 MODE UP DOWN ADJUST BUTTONS CALIBRATION STATUS LED MODE INDICATOR LED S LED DISPLAY LAMP ON STATUS LED ...

Страница 9: ...ser Guide 4 Rear Panel Figure 2 Rear Panel ADDITIONAL I O P2 ANALOG JACK FOOT PEDAL 15 PIN DE STYLE PORTS FOR PLC INTERFACING AC RECEPTACLE AND DUAL FUSE HOLDER D SUB 9 PIN RS 232 PORT P3 STANDARD I O P1 ...

Страница 10: ... produced by the lamp used in this product Always use the UV protective eyewear supplied with the unit and always turn the lamp off before removing lamp housing cover Caution Never look into the light emitting end of the light guide The light could severely damage the cornea and retina of the eye if the light is observed directly Eye shielding must be used at all times as well as clothing to prote...

Страница 11: ... health hazard to humans When unpacking or installing the lamp always wear protective clothing and a face mask Operate lamp only in the S2000 lamp housing This prevents direct viewing of the arc and in the case of lamp bursting contains the lamp particles In the rare instance in which a lamp bursting occurs and the mercury content is released the following safety precautions are recommended all pe...

Страница 12: ...cs will dispose of them in the appropriate manner Caution The lamp module s operational life can be significantly shortened if it is handled incorrectly Do not touch the bulb s glass envelope or the inside surface of the reflector Skin oils can cause the lamp module to fail prematurely Caution Prior to opening the unit and handling the lamp module allow the lamp module to cool down completely appr...

Страница 13: ...nstalled depending on which configuration your S2000 is set 4 1 1 Be sure the AC POWER cord is disconnected from the unit 4 1 2 Remove the screw from the lamp housing side panel using the tool provided and remove the panel from the unit cover Figure 3 Lamp Housing Panel 4 1 3 Carefully remove the lamp module from its container holding only the ceramic component and lamp rim 4 1 4 As illustrated be...

Страница 14: ...in Intelli Lamp sensor connector at the rear of the lamp module and connect it to its mate located on the top of the lamp housing wall CERAMIC COMPONENT LAMP HOLDER LAMP RIM POWER LEADS SPRING CLAMP Tip the Intelli lamp connector will only attach in the correct orientation If you are having difficulty attaching the connector try rotating it by 180º Note if the Intelli lamp connector is not install...

Страница 15: ... as required 4 1 9 Replace the lamp housing side panel and tighten the fastening screw INTELLI LAMP CONNECTOR POWER CONNECTOR CENTER BAR AND LEAD WITHIN THE CUTOUT Note 1 if the lamp housing panel is not secured completely the lamp will not strike and the bulb message will display when power is turned on to the unit Note 2 if the wrong type of lamp has been installed in your unit the message E1 wi...

Страница 16: ...ner and pull out firmly 5 1 5 Refer to section 19 5 for Light Guide Cleaning Instructions Note 3mm single leg liquid light guides are not compatible with the S2000 Tip When the light guide is fully inserted the LED above the light guide port will illuminate green when POWER is on to the unit If the light guide is not fully inserted the LED will illuminate red Note the shutter will not open if the ...

Страница 17: ...for approximately 4 minutes The display will stop flashing when the warm up period has completed Lumen Dynamics recommends 20 minutes to ensure a stable optical output Warning If the lamp is turned off and an attempt is made to turn it back on before it has fully cooled the cool message will appear on the display The lamp will automatically re strike when the lamp has cooled 6 1 6 To power down th...

Страница 18: ...value of irradiance in W cm2 and the front panel CAL LED is ON 7 1 7 Relative mode refers to the condition when S2000 has not been calibrated The display will show the percentage of iris opening and the front panel CAL LED is OFF 7 1 8 Section 8 describes adjusting the light output in both relative and absolute modes in greater details 7 1 9 The S2000 can be calibrated and correspondingly set to a...

Страница 19: ...olute mode Additionally the current light guide diameter value selected by the radiometer will be transferred to the S2000 This light guide diameter will be used by the S2000 to calculate the proper output irradiance in W cm2 7 1 13 For detailed operating instructions regarding calibration of the S2000 refer to the R2000 user s guide RADIOMETER CALIBRATION CABLE S2000 LIGHTGUIDE ...

Страница 20: ...D on the front panel is illuminated The CAL LED indicator will be OFF The display will show CLF and the Closed Loop Feedback LED will be illuminated The only exception is when the unit is set to 100 for maximum output power In this case the Closed Loop Feedback option is not activated Closed Loop Feedback LED will be OFF and the LED display will indicate 100 of iris opening Note The Closed Loop Fe...

Страница 21: ...t intensity even as the lamp ages As a result of lamp aging subsequent exposures may display a higher iris open percentage value however the output intensity will be consistent with the original setting 5 or 200mW cm2 whichever is greater 8 2 Adjusting the Light Output in the Absolute Mode 8 2 1 Set your S2000 to Level Mode by pressing the Mode button until the LEVEL LED on the front display is il...

Страница 22: ...ss the start stop button to close the shutter The open indicator will turn off 8 2 6 To check the percent of iris opening while in the level mode press and hold the Mode button The LED display will indicate XXX percentage of iris opening Note While in Level Mode pressing the start stop button will manually open the shutter until the start stop button is pressed again to close the shutter The timer...

Страница 23: ... Lock the up down buttons A Push and hold the mode button B While pressing the mode button and within 2 seconds press the following sequence of buttons DOWN UP UP UP C After 2 seconds the message LOC will appear for 2 seconds While the system is locked only the mode button and the start stop button will function If the up or down button is pressed the LOC message will appear for 2 seconds 9 1 3 To...

Страница 24: ... up or down button to rapidly increase or decrease the exposure time 10 2 Running a Timed Exposure 10 2 1 While in Timer or Lamp Hours mode press the start stop button The shutter will open and the exposure time will begin to count down When the exposure time has elapsed to zero the shutter will close 10 2 2 Running a timed exposure is only possible when the unit is set in either TIMER or LAMP HOU...

Страница 25: ...en the analog inputs and outputs intensity monitor output and intensity control input are optically isolated from the control circuitry 11 2 Input Output Signals and Descriptions 11 2 1 Output Signals an output signal is said to be active when the output is ON a signal is said to be ON when the transistor side of it s respective optocoupler is conducting current the description of current conducti...

Страница 26: ...User Guide 21 Figure 6 Active Low Signal Circuit Configuration Figure 7 Active High Signal Circuit Configuration ...

Страница 27: ...hoto diode cathode side logic Level active low zero voltage input max open circuit output voltage 6 VDC max current sinking requirement 5mA logic low time for momentary 150mS minimum The diagram below is a sample connection method which may be utilized for the S2000 rear panel I O connections Figure 8 Rear Panel Sample Connection ...

Страница 28: ...4 PLC Signal Descriptions The rear panel has an I O signal connector labelled P1 it is a 15 pin DE style connector It has the following pin outs PIN NO SIGNAL NAME 1 lamp ON output emitter 2 shutter verification output emitter 3 exposure fault output emitter 4 common return for inputs 5 shutter activation start stop input 6 shutter interlock input 7 intensity adjust increase input 8 intensity adju...

Страница 29: ...wing pin out PIN NO SIGNAL NAME 1 alarm clear input 2 shutter alarm out 3 shutter alarm out 4 bulb alarm out 5 bulb alarm out 6 intensity control input 7 intensity monitor output 8 cool out 9 cool out 10 sync out 11 sync out 12 mode button input 13 not used 14 common return for intensity I O 15 common for inputs GND Table 2 P2 Pin Out ...

Страница 30: ...signal duplicates the front panel start stop button When the PLC Shutter signal mode in GUI system administration is selected to Pulsed Mode default a momentary closed contact activates the internal timer When the PLC Shutter signal mode is selected to Level Mode continuous closed contact provides external timing as the iris remains open for the duration of the contact closure Pin 6 Shutter interl...

Страница 31: ...ontact closure type signal This signal must be activated to allow use of the intensity increase or intensity decrease signals as described above 11 2 6 I O signal connector P2 signal descriptions Pin 1 Alarm clear input This signal allows the user to clear the alarm state of the unit Accompanied with an alarm signal would be an audible alarm In addition to clearing the output alarm signal this inp...

Страница 32: ...en 1VDC and 5VDC depending on the sampled intensity The relationship between the intensity and the output voltage is as follows Output Voltage 1 POWER 0 36 W 1W 1 360V 0W 1VDC Pins 8 9 Cool output This signal is enabled when the Cool message is displayed on the LED display Lamp is too hot to strike Once the lamp re strikes the signal disables Pins 10 11 Sync output This signal is enabled only duri...

Страница 33: ...hat is supplied with the unit Plugging the cable into any free serial port of a PC allows data to be transferred to and from a PC to perform a variety of functions All functions are initiated by a program that executes on the PC See Section 16 for further details PIN NO SIGNAL NAME 1 N C 2 RS232 transmit 3 RS232 receive 4 DTR Data Terminal Ready 5 RS232 GND 6 N C 7 N C 8 N C 9 N C Table 3 P3 Pin O...

Страница 34: ...dio style jack that can be connected to a foot pedal supplied with each unit or any other electro mechanical triggering device This is an exposure trigger input used to start an exposure The shutter activation input and front panel START STOP button will also trigger the shutter 12 Viewing the Accumulated Lamp Hours 12 1 1 Your S2000 system automatically accumulates the number of hours that the la...

Страница 35: ... on the display indicate that unit is in Intensity Mode and shutter is open The three digit number shown will indicate the percent iris opening See Section 8 Adjusting the Light Output 6 bulb The bulb message advises the user that the bulb has failed to ignite or extinguished after striking If the bulb message appears immediately after power up and is accompanied by an audible beeping this indicat...

Страница 36: ...m temperature 10 LOC The LOC message appears for about 2 seconds when the up down adjustments are initially locked It also appears if the up or down button is pressed while they are locked advising the user that these functions are locked out See Section 9 Locking and Unlocking the up down Adjustment Buttons 11 ULOC The ULOC message appears for about 2 seconds after unlocking the up down adjustmen...

Страница 37: ...d be serviced 17 E5 The E5 message appears on the display when the communication to the intellilamp has failed Turn off power as per section 6 open the lamp access door and verify all connections are secure Restart the system as per section 6 If error re occurs the unit should be serviced 18 E6 The E6 message appears on the display when an internal error has occurred Turn off power as per section ...

Страница 38: ...he lamp is off Open Shutter Status On The shutter is open Off The shutter is closed Timer Mode On The unit is in timer mode LED display indicates exposure time LEVEL Mode On The unit is in intensity mode LED display indicates intensity in mW cm2 when calibrated OR LED display indicates percentage of iris opening when unit is not calibrated L Hrs Mode On The unit is in lamp hours mode LED display i...

Страница 39: ... alarm 1 Press the front panel start stop button Note Alternately depressing the foot pedal or providing a momentary contact closure on the rear panel analog jack will clear an audible alarm 2 The audible alarm can be also cleared remotely From a PC through I O port when the CLR command is sent to the S2000 Through the PLC I O port via P2 pin 1 and pin 15 ...

Страница 40: ...0 and responses from the S2000 include 3 pieces of data the command section a CRC8 and a carriage return r in C code or chr 13 in basic code The CRC8 one byte value is entered as a hex string Please refer to CRC8 routine for details in section 16 7 16 3 2 When the carriage return character CR is received the S2000 will respond with an error message if there is a problem For example Error Message R...

Страница 41: ...R Response from S2000 Received 6 Get unit status Command to S2000 GUS Response from S2000 n n represents a integer number Example If 50XX r is received the unit status is as follows Alarm Off Shutter Home Pass Lamp On Lamp Ready Yes Shutter Close Front Panel locked Calibration No Exposure Fault No 7 Run a timed exposure Command to S2000 RUN Response from S2000 Received 8 Open Shutter Command to S2...

Страница 42: ...configuration is as follows No abuse bit set surface curing with 20 hours lamp 13 Get calibrated lamp hours Command to S2000 CLH Response from S2000 n n represents a integer number in hours Example If 20XX r is responded the unit was calibrated when the lamp hours was 20 Note Need CAL the S2000 need to be calibrated before the calibrated lamp hours can be obtained 14 Get iris Level Command to S200...

Страница 43: ... n represents a integer number Example If 10XX r is responded the software version on the main board will be 1 0 18 Get software versions from I O board Command to S2000 VIO Response from S2000 n n represents a integer number Example If 10XX r is responded the software version on the I O board will be 1 0 19 Get SERIES 2000 serial number Command to S2000 GSN Response from S2000 n n represents a in...

Страница 44: ...ormation is available The S2000 displays LEVEL instead of irradiance 21 Get SERIES 2000 POWER set point Command to S2000 GPW Response from S2000 n n represents a float number in Watts Example If 1 222XX r is responded the S2000 LEVEL set point will be 1 222 W If the unit is in LEVEL mode Need CAL will be the response 22 Get SERIES 2000 irradiance set point Command to S2000 GIR Response from S2000 ...

Страница 45: ...000 n n represents a float number in Watts per square cm If the unit is in LEVEL mode Need CAL will be the response Example If 35 66XX r is responded the S2000 maximum irradiance output will be 35 66 W cm2 25 Set iris Level Command to Series 2000 SILn Note 0 n 100 n must be a integer number Example If SIL20XX r is set the S2000 iris Level will be 20 Response from S2000 Received if the value n is v...

Страница 46: ...e S2000 is running a timed exposure so the value cannot be set 27 Set LEVEL set point Command to S2000 SPWn Note n 1 5mm LG n 25 8mm LG and n is a float number Example If SPW2 55XX r is set the SERIES 2000 LEVEL set point will be 2 55 W Response from S2000 Received if the value n is valid Invalid if the value n is out of range Need CAL the S2000 need to be calibrated before the value can be set EX...

Страница 47: ...n trigger level mode is disabled If n is 1 then trigger level mode is enabled Response from S2000 Received if the command has been accepted Invalid if the value is outside the specified range 16 5 Command Timing Specification 16 5 1 The minimum amount of time required for data to be received by the PC is over 20mS except for any commands related to iris movement which is over 200msec The next comm...

Страница 48: ...Command 3 Calculate CRC8 for the command GLG Screen Cursor crHourGlass Show hourglass cursor Application ProcessMessages while Tries ComPort FlushInBuffer ComPort FlushOutBuffer ComPort PutString Textcom send the command GLG ComPort PutChar Asc Command_CRC 4 send the CRC8 in hex string ComPort PutChar Asc Command_CRC 0xF send the CRC8 in hex string ComPort PutChar r a 0 UnitResponse TimeOut false ...

Страница 49: ...ffFixed 12 3 Screen Cursor crDefault Cmd33 Enabled true Timer1 Enabled true Application ProcessMessages return if TimeOut MessageDlg No response from Unit mtError TMsgDlgButtons mbCancel NULL Screen Cursor crDefault Cmd33 Enabled true Timer1 Enabled true Application ProcessMessages Bit 0 1 2 3 4 5 6 7 Alarm Lamp Shutter Home Lamp Ready Lock Calibration Exposure Fault Status 0 Off Off Close Pass No...

Страница 50: ... char A unsigned char i CRC8 0 reset CRC8 for i 0 i 8 i data loop A Data get first data byte for LoopCntr 0 LoopCntr 8 LoopCntr A 1 8 bit loop if A CRC8 0x01 test bit 0 of OneWire Data XOR CRC8 CRC8 0x18 toggle bits 3 and 4 of CRC8 CRC8 1 rotate right CRC8 1 time CRC8 0x80 set bit 7 of CRC8 else CRC8 1 rotate right CRC8 1 time return CRC8 ...

Страница 51: ...0 unit The Control Panel provides the following features 1 COM port selection 2 Lamp ON OFF control 3 Shutter control 4 Relative intensity setting control 5 Absolute intensity setting control in power 6 Absolute intensity setting control in irradiance 7 View light guide diameter 8 View calibration expiration time 9 Exposure time setting 10 Display lamp hours Display unit status Alarm Shutter posit...

Страница 52: ...1 Turn on the PC to be used with the S2000 system 17 2 2 It is recommended that any other Windows programs currently in use be shut down 17 2 3 Insert the S2000 Control Panel software CD in the CD ROM drive of your PC 17 2 4 Right click your mouse on the Windows Start button and select Explore 17 2 5 Left click on Explore and select the appropriate drive containing the Series 2000 installation CD ...

Страница 53: ... hours is displayed in the test box labelled Lamp Hours H To update the reading to the current value right click your mouse on the adjacent GET control button Lamp On Off The lamp in the S2000 is turned on automatically once the front panel POWER switch on the unit is set to the ON position To remotely turn the lamp off click on the Turn Lamp Off button in the Lamp Control section on the GUI The f...

Страница 54: ...n irradiance W cm2 is displayed in the Max Irradiance text box Click on the GET button to update the reading to the current value This displayed value is the maximum available output intensity from the unit 17 6 3 Method 1 To set the optical output to a specified irradiance value click on the Irradiance Set point UP and DOWN buttons until the desired value is displayed in the irradiance set point ...

Страница 55: ...e the set button has greyed out the value has been entered 17 8 Exposure Settings Exposure Timer 17 8 1 Ensure that the check box next to Exposure Timer Active S has been selected Exposure time can be selected by the following 2 different methods 17 8 2 Method 1 Adjust the exposure time by clicking on the Up Down buttons located right of the Exposure Timer The exposure time will change in incremen...

Страница 56: ...absolute power W or irradiance W cm2 When the shutter is closed the intensity will read 0 17 11 Shutter Control 17 11 1 To activate the shutter control once the appropriate intensity and exposure settings have been entered click on the Open button in the Shutter Control section 17 11 2 With the Exposure timer active S selected after the exposure cycle has been completed the shutter will close and ...

Страница 57: ...he original alarm condition must be corrected prior to resuming normal operation 17 13 Calibration Status 17 13 1 The S2000 s calibration interval is valid for 112 hours of elapsed lamp on time as measured from the time the unit is calibrated with a R2000 Radiometer The remaining calibration time is displayed in the Calibration In text box in hours To retrieve the current remaining time click on t...

Страница 58: ...en the GUI is in absolute mode the intensity shown changes from relative Level to either Irradiance W cm2 or Power W depending on which mode was selected in the main GUI under Menu Refer to the Control Panel for the maximum irradiance or power available 18 1 5 Input the above parameters for each step maximum of 10 steps and then enter the number of StepCure cycles Press Save to save the input para...

Страница 59: ...rom the S2000 hold the up button while pressing the down button twice Calibration will disable current step cure automatically therefore Step cure has to be updated when unit calibration status changes Downloaded Step cure is also disabled if step cure is run directly from the GUI 18 2 2 The clear step cure button in GUI will remove the step cure process from the S2000 unit connected If no step cu...

Страница 60: ...r will automatically close on the next closure of the contacts if it is open if it is closed it will start the next timed exposure 18 3 3 The level start stop mode opens the shutter when the shutter inputs are closed and closes the shutter when the shutter inputs are opened When PLC Shutter signal mode is in Level Start Stop the S2000 must be also in level mode for the shutter to open close with t...

Страница 61: ...p life 5 Replace the air filter found under the front face plate frequently to ensure unrestricted air flow It is recommended as a minimum that the air filter be removed and washed with a mild detergent and water every time the lamp module is replaced 6 When necessary clean the light emitting end of the light guide using an optical cleaning solution 7 Cleaning of unit is not required however if cl...

Страница 62: ...se drawer which is located in the AC inlet module on the rear panel 19 2 2 Turn off the main POWER switch and remove the AC POWER cord from the unit 19 2 3 Gently pull out the drawer with the aid of a flat head screwdriver 19 2 4 Carefully lever one end of the blown fuse up from its retaining clip with a small flat head screwdriver and lift it out 19 2 5 Replace the damaged fuse s only with the sa...

Страница 63: ... system To change the bandpass filter the system must be returned to your local Lumen Dynamics Service Centre See section 0 Contact Information to find the nearest service centre 19 5 Light Guide Cleaning Instructions for cleaning the light emitting end output of the light guide Materials Lens Tissue sheets Optical Swabs IPA Isopropyl Alcohol solution Wooden stick eg tongue depressor popsicle stic...

Страница 64: ...tes 2 If required remove light guide output end from the production fixture 3 Remove the light guide from the light guide port of the UV spot curing system 4 Place the protective cap onto the light guide input end while cleaning output end 5 Wipe output end of light guide with non abrasive lint free lens tissue or optical swab saturated with an IPA cleaning solution until the staining has disappea...

Страница 65: ...be serviced CA1 None Off Calibration S2000 unit is doing calibration On Calibration Failure The calibration has failed LOC None Off Adjustment Locked The up down adjustment buttons have been locked No changes can be made to the exposure time or iris setting ULOC None Off Adjustment Unlocked The up down adjustment buttons have been unlocked Changes can be made to the exposure time and iris setting ...

Страница 66: ... of default Pulsed Start Stop Shutter will remain open only for the time till the start stop button or foot pedal is pressed L HRS ON The unit is in lamp hours mode LED display indicates accumulated lamps CAL ON Calibration status indicator The S2000 has been successfully calibrated Unit is in absolute mode Calibration will remain in effect for 112 hours of lamp on time CAL OFF The S2000 requires ...

Страница 67: ...s in the ON position 20 1 2 If the LED display lights and the fan starts but the lamp will not turn on check for the following 1 The LED display indicates the bulb message and the system begins to beep This indicates that no lamp is detected Check if the lamp has been installed correctly Refer to Section 4 Installing the Lamp Module 2 The LED display indicates the bulb message after approximately ...

Страница 68: ...ng the Light Output 2 There are no foreign substances on the emitting end of the light guide 3 There are no bends kinks or other physical damage to the guide Replace the light guide if there is any physical damage 4 The lamp has been installed correctly See section 4 Installing the Lamp Module It may be necessary to replace the lamp or to replace the light guide Contact your Lumen Dynamics sales r...

Страница 69: ... sec and lower 21 1 Lamp Module A Rim of Lamp Reflector B Reflector C Back ceramic Mount D Intelli Lamp Connector E Power Connector Lamp Module Lumen Dynamics 200W Mercury DC Standard or Surface Curing Focal Point of Spot 52 875 mm Lamp Module Life 2000 hours Warm Up 4 minutes min Lamp Voltage new bulb 60VDC nominal Lamp Voltage range 36VDC to 95VDC Lamp Current 3 33A typical 5 7A maximum Warning ...

Страница 70: ...ding they are returned in its original packaging Lumen Dynamics will dispose of them in the appropriate manner Warning Hg LAMP CONTAINS MERCURY Manage in Accord with Disposal Laws see www lamprecycle org or 1 800 668 8752 ...

Страница 71: ...ge 100 240VAC 50 60Hz Current 3 5A max at 120VAC 2 0A max at 240VAC Input Surge 50A max cold start Protection Short circuit auto recovery Overvoltage up to 135 5 of nominal EMI filtering integrated into the power supply Power supply has integrated thermal cut off Fuse Rating Dual fuse system each fuse rated at F5 0A 250V 5x20mm type located in the AC receptacle Note 3mm single leg liquid light gui...

Страница 72: ...n the side of the front panel which is used to connect to the R2000 Radiometer for calibration of the S2000 21 4 USB Interface 21 4 1 A USB interface an optional feature can be provided through the existing RS 232 port via an external RS 232 to USB converter device A RS 232 to USB converter device such as part number RS232B by FTDI Future Technology Devices International Inc or equivalent can be u...

Страница 73: ...ps xenon arc lamps switch mode POWER supplies pulsed lasers x ray equipment welding equipment and RFI generators are a few classic examples of large noise producing equipment 21 6 2 Determining the Noise Level in Your Environment If the S2000 along with other pieces of equipment operate abnormally in an intermittent or continuous manner it is quite possible that large Levels of noise are present I...

Страница 74: ... Suitable ferrite shield beads are P N 0443164251 by Fair Rite Products Corp or P N 28A2025 0A0 by Steward Noise suppression products are also available in the form of AC power bars that incorporate surge suppression and noise suppression circuitry Either type will help in protecting the S2000 from noise Providing the S2000 with an AC line that is separate from any other noise producing equipment ...

Страница 75: ...se Part 1 General Requirements Electrical Equipment for Laboratory Use Part 1 General Requirements IEC Equipment Class I Installation Category II Pollution Degree 2 Electromagnetic Compatibility EN 61326 1 2006 Electromagnetic Compatibility Immunity Testing Measurement Control and Laboratory Equipment EN55011 2007 CISPR11 2004 Class B Group 1 Emissions Testing Industrial scientific and Medical Equ...

Страница 76: ...ul interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the r...

Страница 77: ...hina RoHS regulations 有毒有害物质名称及含量的标识格式 部件名称 有毒有害物质或元素 铅 Pb 汞 Hg 镉 Cd 六价铬 Cr6 多溴联苯 PBB 多溴二苯醚 PBDE 弧光灯模块 200W x x 点火模块 压装 缓冲器印刷电路板组件 灯管镇流器印刷线路板组件 电源印刷电路板组件 光圈印刷电路板 显示器 前面板印刷电路板组件 输入 输出信号印刷电路板组件 表示该有毒有害物质在该部件所有均质材料中的含量均在SJ T11363 2006标准规定的限量要求以下 x 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出SJ T11363 2006标准规定的限量要求 企业可在此处 根据实际情况对上表中打 的技术原因进行进一步说明 ...

Страница 78: ...those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Those systems will reuse or recycle most of the materials of your end life equipment in a sound way 22 3 4 The crossed out wheeled bin symbol indicated above invites you to use those systems If you need more information on the collection reuse and recy...

Страница 79: ...f reporting to a an authorized Lumen Dynamics Service Centre Lumen Dynamics will repair or replace these reported defects free of charge The equipment must be sent postage and carriage paid 23 1 5 Package the equipment in its original shipping case or as appropriate to prevent damage during transport 23 1 6 In the case of damage caused by wear and tear careless handling neglect by the use of force...

Страница 80: ...rventions and repairs not carried out by an Lumen Dynamics Service Centre to the S2000 system the guarantee ceases to be valid 23 3 Returning your S2000 to Lumen Dynamics 23 3 1 Please make a note of the problem encountered the steps followed to isolate the problem and the result of any trouble shooting steps taken 23 3 2 Telephone the nearest Lumen Dynamics Service Centre to obtain a Return Autho...

Страница 81: ...ics 2260 Argentia Road Mississauga Ontario L5N 6H7 CANADA Tel 1 905 821 2600 Toll 1 800 668 8752 USA and Canada Fax 1 905 821 2055 Web site www ldgi com http www ldgi omnicure com contact php Sales omnicure ldgi com Tech Support techsupport ldgi com ...

Страница 82: ...wing modifications Section 11 2 5 The Rear Panel I O signal connector labelled P1 PIN 5 shutter activation Section 20 Front Panel LED Descriptions level indicator Section 16 4 Command Descriptions New PLC command features Section 18 1 5 Stepcure functionality Section 18 3 PLC Shutter Signal Mode ...

Отзывы: