User Guide
- 37
–
10. Turn On Lamp
Command to S2000;
TON
Response from S2000;
Received
11. Turn Off Lamp
Command to S2000;
TOF
Response from S2000;
Received
12. Get Lamp Configuration
Command to S2000;
GLH
Response from S2000;
n
(n represents a integer number)
Bit 15: Abuse Bit (0=OK, 1=lamp abused)
Bit 14: Lamp Type (0= surface curing, 1=standard)
Bit 13
– Bit 0: Lamp Hours
Example:
If “20XX\r” is responded, the lamp configuration is as
follows:
No abuse bit set, surface curing with 20 hours lamp.
13. Get calibrated lamp hours
Command to S2000;
CLH
Response from S2000;
n
(n represents a integer number in hours)
Example:
If “20XX\r” is responded, the unit was calibrated when
the lamp hours was 20.
Note:
Need CAL
– the S2000 need to be calibrated before the
calibrated lamp hours can be obtained.
14. Get iris Level
Command to S2000;
GIL
Response from S2000;
n
(n represents a integer number 1-100)
Example:
If “20XX\r” is responded, the iris Level will be 20%.
15. Clear unit calibration
Command to S2000;
CLC
Response from S2000;
“Done”