User Guide
- 26
–
Pin 12: Lamp off/on input
This contact closure type signal toggles power to the lamp. If the lamp is OFF,
a momentary closed contact will turn the lamp on. Conversely, if the lamp is
ON, a momentary closed contact on this input will turn the lamp off.
Pin 13: Shutter verification mode input
Providing a contact closure changes the mode of the shutter verification output
signal found on pins 2 and 10 to a shutter position indicator. When this feature
is in use, pins 2 and 10 will be active when the shutter is open and inactive
when the
shutter is closed.
Pin 14: Front Panel lock/ unlock input
This contact closure type signal locks/ unlocks the front panel controls.
Pin 15: Intensity lock/ unlock input
The intensity lock/ unlock input is a contact closure type signal. This signal
must be activated to allow use of the intensity increase or intensity decrease
signals as described above.
11.2.6
I/O signal connector (P2): signal descriptions
Pin 1: Alarm clear input
This signal allows the user to clear the alarm state of the unit. Accompanied
with an alarm signal would be an audible alarm. In addition to clearing the
output alarm signal this input signal would allow the user to shut down the
audible alarm.
When an activation signal is received, the alarm state of the S2000 will be
cleared.
Pins 2 (-) & 3 (+): Shutter alarm output
This signal provides an indication of a shutter failure. The output signal is ON
when an alarm condition is present. The optocoupler is OFF at all other times.
Pins 4 (-) and 5 (+): Bulb alarm out
This signal provides an indication of a lamp failure. The output signal,
accessible through pins 4 and 5, is ON when an alarm condition is present. The
optocoupler is OFF at all other times.