User Guide
- 8
–
Caution!
The lamp module’s operational life can be significantly shortened if handled
incorrectly. Be sure only to handle the ceramic surfaces and the lamp rim.
Do not touch the bulb’s glass envelope or the inside surface of the reflector.
Skin oils can cause the lamp module to fail prematurely.
4 Installing the Lamp Module
The S2000 curing system is available configured in 2 different versions, one version
is for standard curing and one version is for surface curing. The appropriate type of
lamp must be installed depending on which configuration your S2000 is set.
4.1.1
Be sure the AC POWER cord is disconnected from the unit.
4.1.2
Remove the screw from the lamp housing side panel using the tool provided
and remove the panel from the unit cover.
Figure 3 Lamp Housing Panel
4.1.3
Carefully remove the lamp module from its container, holding only the ceramic
component and lamp rim.
4.1.4
As illustrated below position the lamp facing towards the front of the unit with
the POWER leads facing towards you. The lamp should be aligned so that the
leading edge of the reflector (lamp rim) fits into the mounting groove on the
lamp holder assembly.
LAMP HOUSING
SIDE PANEL
Note: Refer to Section 3
– Safety Precautions before proceeding