![Lumen Dynamics Group OmniCure S2000 Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/lumen-dynamics-group/omnicure-s2000/omnicure-s2000_user-manual_1947998011.webp)
User Guide
- 6
–
Danger
This unit contains HIGH VOLTAGE components. It is recommended that ONLY
QUALIFIED TECHNICAL PERSONNEL perform any testing or repairs.
Monitoring the unit during manual operation
The Level of UV energy supplied by the S2000 is sufficient to ignite flammable
substances. During manual operation, the unit must be attended at all times by a
qualified operator. The unit must not be left unattended while turned on. If an
operator leaves the work area of the unit, the POWER switch must be turned off.
Monitoring the unit during automated operation
The Level of UV energy supplied by the S2000 is sufficient to ignite flammable
substances. Therefore, when the unit is operated unattended in an automated
environment, an alarm function must be provided by the user to indicate a
malfunction in the associated equipment used.
Warning
Hg
– LAMP CONTAINS MERCURY, Manage in Accord with Disposal Laws, see:
www.lamprecycle.org
or 1-800-668-8752
Danger: Exposure to Mercury represents a health hazard to humans.
When unpacking or installing the lamp, always wear protective clothing and a face
mask. Operate lamp only in the S2000 lamp housing. This prevents direct viewing
of the arc and in the case of lamp bursting, contains the lamp particles. In the rare
instance in which a lamp bursting occurs, and the mercury content is released, the
following safety precautions are recommended: all personnel should be
immediately evacuated from the area to prevent inhalation of the mercury vapour.
The area should be well ventilated for a minimum of 30 minutes. Prior to clean up
ensure an approved mercury respirator mask and non-porous gloves such as latex
or rubber are used. After the lamp housing elements have cooled, the mercury
residue should be collected with the use of a special absorbing agent available
from laboratory equipment suppliers.
Listed below are examples of internet web sites for obtaining Mercury Spill Kits;
http://www.rosshealthcare.org/Mercon.spill.kits.htm
http://www.coleparmer.ca/catalog/product_index.asp?cls=43577
http://www.environmental-
expert.com/technology.aspx?idCategory=2054&word=mercury%20spill%20kits
http://www.alibaba.com/products/spill_kit/4.html